GPIB Troubleshooting Guide

If you have difficulty operating the 6810B via GPIB, refer to the symptoms below:
Symptom
No response from 6810B upon a GPIB command
(remote light does not come on)
Slow or unexpected response to GPIB
commands
Device does not respond to command (remote
light is on)
Bus hangs
01_10
T R O U B L E S H O O T I N G
GPIB Troubleshooting Guide
Corrective Action
Make sure you purchased Option 1268 with your
6810B. A connector on the rear panel does not
necessarily mean the device is GPIB capable.
Check that a GPIB cable is connected between the
6810B and the system controller (your computer). This
cable should be less than 3 meters long.
Check that the GPIB address is set properly and the
controlling software is sending commands to the proper
address.
Make sure no two devices have the same GPIB
address.
Check that there are fewer than 15 devices on the bus
and there is less than 20 meters of total cable.
Check that the GPIB controller card in the host
computer is configured properly. See Appendix C.
Try isolating the 6810B by removing all other
instruments from the bus.
Make sure no two devices have the same GPIB
address
Check that there are fewer than 15 devices on the bus
and there is less than 20 meters of total cable.
Make sure the GPIB controller card in the host
computer is configured properly. See Appendix C.
Isolate the 6810B by removing all other instruments
from the bus.
Read the error queue remotely (Err?). The command
syntax or command structure may be in error.
Read the status byte (*STB?) and condition register
(COND?) for possible device problems.
Make sure the GPIB controller card in the host
computer is configured properly. See Appendix C.
Place a software timeout around the command in
question to prevent the hang.
Check the syntax of the command that is causing the
hangup.
Make sure the controlling software is not requesting
information from the 6810B without first sending a
query.
Make sure the GPIB controller card in the host
computer is configured properly. See Appendix C.
Check GPIB cables and the GPIB address.