Symptoms
Does not power up.
ZERO function does not work.
Display shows "− OL −".
Display shows "− LO −".
68
FPM-8200
T R O U B L E S H O O T I N G A N D M A I N T E N A N C E
Troubleshooting Guide
Corrective Actions
Check the power cord to make sure that it is properly connected. To
check a power connection, connect the cord to another known
working instrument.
Error message E−531 normally indicates there is too much room light
for a proper ZERO. The optic input connector must have an adaptor
with dark fiber attached.
Connector adaptor: Make sure you have a patchcord connected.
Leave the other end of the patchcord connected to your source with
its output set OFF.
Bare fiber adaptor ring: Make sure the ILX BF−820 Bare Fiber Holder
is installed, with a fiber. Leave the other end of the fiber connected to
your source with its output set OFF.
Agilent 81000BA bare fiber holder: Make sure Agilent holder is
installed, including a fiber. Leave the other end of the fiber connected
to your source with its output set OFF. Turn the fiber holder so the
large V−groove along the top is facing down, or away from room light.
If the error persists, contact ILX Lightwave.
(OverLoad)˙Too much optical input power for this range. Press UP
on the front panel to change the gain range. Or select AUTO/MAN
so the AUTO indicator is ON to automatically select an appropriate
gain range. You can then select AUTO/MAN again (AUTO indicator
OFF) if you want to lock it manually into one range.
If you see an − OL − indication in auto−range mode, or in the highest
range (lowest gain), the optical input power is beyond the FPM−8200
upper limit.
If you see an − OL − indication in manual−range mode, then the
optical input power is greater than 99.1% of full scale for the current
gain range.
(LOw)˙Not enough optical input power for this range. Press DOWN
on the front panel to change the gain range. Or select AUTO/MAN
so the AUTO indicator is ON to automatically select an appropriate
gain range. You can then select AUTO/MAN again (AUTO indicator
OFF) if you want to lock it manually into one range.
If you see an − LO − indication in auto−range mode, or in the lowest
range (highest gain), the optical input power is below the FPM−8200
lower limit.
If you see an − LO − indication in manual−range mode, then the
optical input power is less than 5% of full scale for the current gain
range.