Common Commands and Queries

This section contains a list of the common commands and queries which are
supported by the 6810B. These commands are distinguished by the * (asterisk)
which begin each mnemonic. The common commands and queries are listed in
alphabetical order, and a brief description of their functions and syntax is given
below. For more information on these commands, refer to the ANSI/IEEE 488.2-
1987 standards reference.
*CAL?
Response
*CLS
01_10
G P I B / I E E E 4 8 8 . 2 R E M O T E O P E R A T I O N
This command is the CALIBRATION QUERY. The analog to digital (A/D) converters
within the 6810B require periodic calibration to maintain specified accuracy. This
calibration is performed automatically by the internal microprocessor once every ten
minutes. The calibration query can be sent to execute the calibration independent of the
routine ten minute cycle.
When.sent, the 6810B performs the A/D calibration. The query responds with a message
indicating that calibration has been completed successfully.
*
:
0 - Successful calibration
1 - Problem encountered during calibration
Sending the CLEAR STATUS command resets the Standard Event Register (*ESR?),
Status Byte (*STB?) and Error Queue (ERR?) to zero. This command may be used to
clear the status registers before enabling SRQ generation from instrument events.
*
:
Common Commands and Queries
?
CAL
CLS
3