View Reference; Brightness Control; Zoom Control; Rate Control - ILX Lightwave OMM-6810B User Manual

Optical power and wavelength meter
Table of Contents

Advertisement

View Reference

When the DISPLAY VIEW REF button is pressed, the reference values are shown
in the units of the present display. The reference values are shown for three
seconds or until the DISPLAY VIEW REF button is pressed again. If the DISPLAY
VIEW REF button is held down, the reference values stay on the display.

Brightness Control

Zoom Control

Pressing DISPLAY ZOOM again places the bargraph into normal sensitivity mode
where each segment represents 7% of the total power within the selected gain
range. The end points of the bargraph are solidly illuminated.

Rate Control

The measurement rate for the SLOW rate is 60ms/Sample and the input
bandwidth is 1 Hz. A sliding average is taken with sixteen samples. The display
is updated every 60ms and the rate indicator flashes at a frequency of 0.5 Hz.
01_10
I N S T A L L A T I O N A N D O P E R A T I O N
Pressing the DISPLAY VIEW REF button displays the power
and wavelength reference points on the appropriate
display(s). These references are set via the DISPLAY
Δ
and
buttons.
The DISPLAY BRIGHT button is used to adjust the intensity
of the front panel display. Repeatedly pressing this button
cycles the brightness through five possible intensity steps
ranging from brightest to dark. The keypad is back lighted
for viewing in a dark environment. The brightness control
also affects the illumination of the switches.
The DISPLAY ZOOM button controls the sensitivity of the
bargraph display. Pressing DISPLAY ZOOM forces the
bargraph into a high sensitivity mode where each segment
represents 0.7% (ten times the normal sensitivity) of the total
power within the selected gain range. In ZOOM mode, the
end points of the bargraph blink.
The DISPLAY RATE button controls the speed at which
measurements are taken by the 6810B. Repeatedly
pressing DISPLAY RATE causes the instrument to cycle
through the three possible rates. The measurement rate is
determined by viewing the flashing indicator above the
POWER display.
Front Panel Controls
OMM-6810B
2