Table 6-2: Model 1140A Error Codes - Ectron 1140A Instruction Manual

Thermocouple simulator-calibrator
Table of Contents

Advertisement

Remote Operation
Query the Unit's Memory Files :MEMory:CATalog?
Returns a comma-separated list of file names.
Example: :MEM:CAT? might return TEST1,TEST2.
Code
Description Returned
0 No error
101 Temperature out of range
102 Reference junction temp
out of range
103 Temp out of range for
thermocouple
104 Ref jct out of range for
thermocouple
105 Voltage out of range
106 Output overload
107 Thermocouple invalid for
temperature scale
–100 Command error
–101 Command not allowed in
Standby mode
–315 Nonvolatile memory error
–350 Queue overflow
Query the First Error in the Queue :SYSTem:ERRor?
This command pops the oldest error out of the queue and returns its numeric code as well as
a brief description, separated by a comma. Refer to
Example: :SYST:ERR? might return −100,Command error.
Query All Errors in the Queue :SYSTem:ERRor:ALL?
Example: :SYST:ERR:ALL? might return:
−100, Command error
105,Voltage out of range
etc.

Table 6-2: Model 1140A Error Codes

The simulation temperature being programmed is out of
range for the current thermocouple type.
The reference-junction temperature being programmed is
out of range for the current thermocouple type.
The current reference-junction temperature is out of range
for the thermocouple type being programmed.
The current simulation temperature is out of range for the
thermocouple type being programmed.
The voltage being programmed is out of range.
The Model 1140A is incapable of supplying the
programmed output.
The thermocouple type commanded is not available in the
current temperature scale.
Unrecognized query, command, or command parameter.
A command was received which could not be acted on in
Standby mode.
An internal memory error occurred at startup.
An overflow occurred in the error queue; the older error(s)
have been discarded.
6-10