Calibration Procedure
PROCEDURE
Preliminary Setup
1.
Turn on the Model 1140A.
2.
Restore the Model 1140A to its default settings.
a.
Press MENU to go the MAIN MENU. The cursor must be absent. If the cursor is
present, press ESC and then MENU.
Scroll to MAINTENANCE and press ENTER to go the MAINTENANCE MENU.
b.
Scroll to RESET INSTRUMENT TO DEFAULT VALUES and press ENTER.
c.
Scroll to YES and press ENTER.
d.
Press ESC to return the MAIN MENU.
e.
Linear-voltage-tests Setup
Program the Model 1140A as follows:
Set OUTPUT MODE to VOLTAGE.
1.
Scroll to INSTRUMENT MODE and press ENTER to go to the INSTRUMENT MODE
a.
MENU.
Scroll to OUTPUT MODE and press ENTER.
b.
Select VOLTAGE and press ENTER.
c.
Press ESC to return to the MAIN MENU.
d.
Set MATERIAL to COPPER.
2.
At the MAIN MENU, scroll to OUTPUT and press ENTER to go to the OUTPUT
a.
MENU.
Scroll to MATERIAL and press ENTER.
b.
Select COPPER and press ENTER.
c.
Set TERMINALS to BINDING POSTS.
3.
At the OUTPUT MENU, scroll to TERMINALS and press ENTER.
a.
Select BINDING POSTS and press ENTER.
b.
Press MENU to return to the operating screen.
c.
d.
The operating screen should be as is shown in Figure 11-1.
Figure 11-1: Initial Operating Screen for Voltage Tests