THERMOCOUPLE-ACCURACY TABLES

The following tables are provided to easily ascertain the maximum error allowed for the
thermocouple types supported. They apply when either sourcing or measuring temperature with
thermocouple wires. A 95% confidence level (k = 2) is assumed. The errors were derived using
the RSS (root of the sum of the squares) of all the applicable sources for error. These errors
include:
1.
Cold-junction compensation (if used without CJC, errors are reduced).
Variation of the environmental temperature by ±5°C from calibration temperature. If
2.
a better temperature environment is maintained, errors will be reduced.
3.
Noise, 1 µV peak in a "0.1 Hz to 10 Hz" bandwidth.
4.
Accuracy, as listed on page
Uncertainty limits (k = 2) of Ectron's temperature measurement system, ±0.01°C.
5.
Conformity, ±0.4 µV.
6.
LCD with adjustable contrast and backlight, which can be
turned off, on, or timed out with settings of 30 seconds,
one minute, two minutes, or five minutes.
USB included; Ethernet, GPIB, and RS-232 optional.
Carrying case.
Calibration kit consisting of a terminal cover, shorting bar,
low-thermal cable, calibrated Type T thermocouple, and
calibrated Type E thermocouple.
0°C to +50°C.
−20°C to +60°C.
0°C to +60°C.
+5°C to +45°C.
368 mm (14.50") wide.
381 mm (15.00") deep.
102 mm (4.00") high, including feet.
482 mm (19.00") wide.
396 mm (15.60") deep.
89 mm (3.50") high, without feet.
10% to 90% noncondensing.
4.5 kg (10 lb) without battery; 5.8 kg (13 lb) with battery.
2-1
for periods of 30 days, six months, and one year.
2-3
Specifications