Data Logging Menu And Downloading Data; Figure 4-9: Data Logging Menu - Ectron 1140A Instruction Manual

Thermocouple simulator-calibrator
Table of Contents

Advertisement

Operation

Data Logging Menu and Downloading Data

In the METER mode, the Model 1140A has the capability to log, store, display, and download
10,000 data points, each of which contains a date, time, and the data itself. By going to the
DATA LOGGING MENU, the user can set the time interval between data points and start logging
data. Once data has been stored, the user can view the data on-screen and download the data in
various formats.
Initial Setup
Begin data logging by setting all front-panel parameters to the desired settings. If in the
TEMPERATURE mode, these include:
Thermocouple type
Thermocouple-offset units (if using alloy connections)
Thermocouple offset (if using alloy connections)
Reference-junction-temperature units
Reference-junction temperature
Instrument mode (must be METER)
Output mode (TEMPERATURE)
System-temperature units
Material
Terminals used (if using ALLOY connections)
Copper offset units (if using COPPER connections)
Copper voltage offset (if using COPPER connections)
Autozero enabled
Autozero temperature offset (if autozero is ON)
If in the VOLTAGE mode, these include:
Instrument mode (must be METER)
Output mode (VOLTAGE)
System voltage units
Material (COPPER)
Copper offset units
Copper voltage offset
Autozero enabled
Autozero voltage offset (if autozero is ON)
To facilitate setup, once the front-panel settings are complete, the user may want to store the
settings as a memory file for later recall using the MEMORY MENU.
DATA LOGGING MENU
START LOGGING NEW DATA
LOGGING INTERVAL
VIEW LOGGING SETUP
VIEW LOGGED DATA (3456 ENTRIES)

Figure 4-9: Data Logging Menu

4-16

Advertisement

Table of Contents
loading

Table of Contents