Maintenance Menu

Use the MAINTENANCE MENU to SET DATE, SET TIME, and RESET INSTRUMENT TO DEFAULT
VALUES. The current settings for date and time are displayed above the menu options.
ALIGNMENT will not appear unless the alignment switch is "on."
Set Date
Selecting SET DATE takes the user to a 100-year calendar (January, 2000 through December,
2099) from which the user can easily set the date using the ARROW KEYS or the ENCODER.
Set Time
The hours (0 to 23), minutes, and seconds can be set using ARROW KEYS or the ARROW KEYS
with the ENCODER. The KEYPAD is not active to set time.
Alignment
When the alignment-enable switch (recessed in a square hole in the bottom cover just to the rear
of the KEYPAD) is switched to the user's right, the ALIGNMENT sub-menu can be accessed. For
alignment instructions, see
Power-on Standby Mode
This option controls whether the Model 1140A will be in standby mode when first turned on. The
default setting is ON. If changed to OFF, the instrument will not enter standby mode on power-up.
In either case, the OPR/STBY (ENTER) key still toggles standby off or on.
Reset Instrument to Default Values
Selecting this option and then selecting YES with confirmation to reset all settings will reset the
Model 1140A to those settings in Table 4-2. ALIGNMENT of the instrument, MEMORY and
SEQUENCE settings, and settings from the DISPLAY MENU and MAINTENANCE MENU are not
affected.
MAINTENANCE MENU
10-MAY-2017
15:31:14
SET DATE
SET TIME
ALIGNMENT
POWER-ON STANDBY MODE
RESET INSTRUMENT TO DEFAULT VALUES

Figure 4-13: Maintenance Menu

Section
X, Alignment.
4-21
Operation