Replace a defective fuse only with a fuse of the original rating.
Never short-circuit fuse or fuse holding.
Disconnect test leads or probe from the measuring circuit before
switching modes or functions.
Do not conduct voltage measurements with the test leads
connected to the µA/mA/A- and COM-terminal of the equipment.
The 10A-range is protected by fuse 10A/1000V.
To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any resistance
measurements.
Do not conduct current measurements with the leads connected
to the V/Ω-terminals of the equipment.
Check test leads and probes for faulty insulation or bare wires
before connection to the equipment.
Please use only 4mm-safety test leads to ensure immaculate
function.
To avoid electric shock, do not operate this product in wet or
damp conditions. Conduct measuring works only in dry clothing
and rubber shoes, i. e. on isolating mats.
Never touch the tips of the test leads or probe.
Comply with the warning labels and other info on the equipment.
The measurement instrument is not to be to operated
unattended.
Always start with the highest measuring range when measuring
unknown values.
Do not subject the equipment to direct sunlight or extreme
temperatures, humidity or dampness.
Do not subject the equipment to shocks or strong vibrations.
Do not operate the equipment near strong magnetic fields
(motors, transformers etc.).
Keep Allow the equipment to stabilize at room temperature
before
taking
measurements).
hot soldering irons or guns away from the equipment.
up
measurement
-42-
(important
for
exact