Extech Instruments EC210 User Manual

Conductivity meter

Advertisement

Quick Links

User Guide
 
 
Conductivity Meter 
 
Model EC210 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Advertisement

Table of Contents
loading

Summary of Contents for Extech Instruments EC210

  • Page 1 User Guide     Conductivity Meter    Model EC210                                                                     ...
  • Page 2 Introduction  Thank you for selecting the Extech EC210 conductivity meter which simultaneously displays  conductivity and Temperature. Conductivity is measured with a remote electrode that includes a  thermistor for measuring temperature. This device is shipped fully tested and calibrated and, with  proper use, will provide years of reliable service. Please visit our website (www.extech.com) to  check for the latest version of this User Guide.    Features   Three ranges 2000uS, 20mS, and 100mS with auto range functionality   Conductivity, TDS and temperature measurement functions   Detachable probe for ease of use in a variety of measurement environments   User adjustable automatic temperature compensation (set to ‘0’ for uncompensated  Conductivity measurements)   Carbon rod electrode type for longevity   LCD display shows Conductivity and Temperature simultaneously   DATA HOLD for holding displayed reading   Record and view MIN MAX readings   Programming mode for selecting  F temperature units, enabling/disabling the auto  power off utility, and programming the temperature compensation functionality    EC210-en-GB_V1.1 11/14...
  • Page 3: Meter Description

    Meter Description    1. Main measurement display  2. Temperature reading  3. Power/Escape button  4. Function/Hold/Up arrow button  5. Record/Enter button  6. Setting/down arrow button  7. Battery compartment (rear)  8. Probe connection  9. Probe        EC210-en-GB_V1.1 11/14...
  • Page 4: Conductivity Calibration

    Conductivity Calibration  Connect the Conductivity electrode to the input socket at the bottom of the meter.  Switch the meter ON by pressing the POWER/ESC button.  Prepare a Conductivity Standard Solution (for example, 1.413uS).  Hold the probe by its handle and immerse the sensing electrode completely in the  standard solution. Shake/stir the sensing electrode to allow the electrode’s internal air  bubbles to escape from the sensing electrode.  Press the HOLD button and the display will show the HOLD icon.  Press the REC button and the upper display area will show CAL and the lower display area  will show YES. To abort at this point, simply wait and the meter will revert to the normal  operating mode.  To continue, press the ENTER button and the display will show the Conductivity reading,  the measurement unit, and, shortly after, the flashing CAL icon. Note that the meter will  show END if the electrode does not detect an appropriate solution value.  While the CAL icon is flashing use the arrow buttons to adjust the displayed Conductivity  reading to match the Conductivity value of the solution.  After a few seconds the meter will display END. Calibration is complete.  Conductivity Measurements  Measurement Preparation  Before taking measurements, perform the calibration described above if necessary.  Taking Conductivity Measurements  Connect the electrode to the probe input socket.  Switch the meter ON by pressing the POWER/ESC button.  If the display does not show Conductivity units (uS or mS), press and hold the FUNCTION  button until the Conductivity units are displayed.  The lower display will indicate the temperature value.  The upper display shows the conductivity value.   Hold the electrode in hand and fully immerse the sensing head in the standard solution.  The upper display will show the conductivity value of the measured solution. The bottom  display will show the temperature value of the measured solution.  Rinse the probe with distilled water after each use.    EC210-en-GB_V1.1 11/14...
  • Page 5 TDS Measurements  Connect the electrode to the probe input socket at the bottom of the meter. Immerse the  probe in the measurement solution.  Switch the meter ON by pressing the POWER/ESC button. Press and hold the FUNCTION  button until the ppm units are displayed.  The lower display shows the temperature value.  The upper display indicates the TDS value (measured in ppm).   Hold the electrode in hand and fully immerse the sensing head in the standard solution  The upper display will show the ppm value of the measured solution. The bottom display  will show the temperature value of the measured solution.  Rinse the probe with distilled water after each use.  Press and hold the FUNCTION button for at least 2 seconds to return to the conductivity  measurement mode.    Solution Temperature  The lower display indicates the temperature of the measured solution. Refer to the Setup Mode  section for instructions on selecting  F temperature units and for adjusting the automatic  temperature compensation.    Data Hold  Press the Function/Hold key to freeze the displayed value. The LCD will display HOLD along with  the held reading. Press the Hold key again to release the data hold function.    MIN‐MAX Data Recording  When selected, the Data Recording function records the Min and Max readings:  Press the REC button once. The REC indicator will appear on the display. Press REC button  once again and REC MAX will display on the screen along with the maximum value.  Press REC again and the display will show REC MIN and the minimum value will be  displayed.  To release the held MAX or MIN reading, press the HOLD button while either the REC MAX  or REC MIN icons are visible; now only the REC icon will be visible. The meter is now  continuing to monitor the highest (MAX) and lowest (MIN) readings but is displaying real  time measurements. To view the MAX and MIN values again use the REC button as  described earlier.  To exit the Record mode, press and hold the REC button for at least 2 seconds. The display  will return to the normal operating mode.    EC210-en-GB_V1.1 11/14...
  • Page 6: Setup Mode

    Setup mode  Note: If the HOLD or MIN/MAX functions are active the Setup mode cannot be accessed.     To enter the Setup Mode press and hold the SETTING button for at least 5 seconds (until the beep  is heard) and then use the SETTING button to scroll through the setup options. The options are:    SEt Automatic temperature compensation percentage; settable from 0.00 to 5.00% per °C  (the default is 2.00%). Set to 0.00% to take uncompensated Conductivity measurements.  When power is cycled this parameter will revert to the default 2.00%    °C/F Select the temperature units (°C °F)    OFF/ON Auto power off disable/enable    Continue with the instructions below:    Temperature Compensation Percent Setting (SEt)  When the display shows SEt press ENTER. SEt will flash and the upper display will show the  temperature compensation percent. Use the ▲ and ▼ buttons to set the desired  temperature compensation percentage. After selecting the desired value, press ENTER to  save the data. The display now shows the temperature units. Continue below:  Temperature Units Select  The lower display shows C or F. Use the ▲ button to select the desired unit of measure and  then press ENTER to save. The lower display should now show OFF YES or NO. Continue  below:  Auto Power OFF Enable/Disable  The display will show OFF YES or OFF NO. Use the ▲ button to select YES or NO.  YES: Auto power OFF is enabled (meter switches off automatically after 10 minutes)  NO: Auto power OFF is disabled  Press ENTER to save the selection. The meter will return to the normal operating mode.    EC210-en-GB_V1.1 11/14...
  • Page 7: Battery Replacement

    Battery Replacement  The 9V battery that powers the instrument requires replacing when the battery icon flashes on  the display. To replace the battery follow the steps below:  Disconnect the probe from the instrument.  Open the rear battery compartment using a screwdriver to remove the two Phillips head  screws that secure the compartment.  Carefully remove the old battery from the compartment lid chamber and gently unsnap  the battery’s terminals from the wired (red/black) connectors.  Install a new 9V battery by snapping the battery terminals onto the wired connectors  (observing correct polarity) and inserting the battery into the compartment lid chamber.  Install the compartment lid chamber into the meter housing and secure with the two  screws.    Never dispose of used batteries or rechargeable batteries in household waste.  As consumers, users are legally required to take used batteries to appropriate  collection sites, the retail store where the batteries were purchased, or wherever  batteries are sold.  Disposal: Do not dispose of this instrument in household waste. The user is obligated to take  end‐of‐life devices to a designated collection point for the disposal of electrical and electronic  equipment.       EC210-en-GB_V1.1 11/14...
  • Page 8: General Specifications

    Power Consumption  Approx. 6.0 mA DC  Weight  295 g (0.65 lbs.) including batteries & probe  Dimensions  Meter: 135 x 60 x 33mm (5.3 x 2.4 x1.3")  Probe: 22mm (0.87”) diameter x 120mm (4.72”) length      Electrical Specifications  Measurement  Range  Resolution  Accuracy   Conductivity  20.00mS/100.0mS/2000 uS  0.01mS/0.1mS/1uS  + (2%FS+1dgt)  TDS  1,200/12,000/66,000 ppm  1/10/100ppm  + (2%FS+1dgt)  Automatic  0 to 50 C (32 to 122  F)  Temperature  Compensation  range        Copyright © 2014 FLIR Systems, Inc.  All rights reserved including the right of reproduction in whole or in part in any form  www.extech.com         EC210-en-GB_V1.1 11/14...