Introduction Thank you for selecting the Extech EC210 conductivity meter. The EC210 simultaneously indicates conductivity and temperature readings. conductivity is measured with a remote electrode which includes a thermistor temperature sensor. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Conductivity Calibration Connect the conductivity electrode to the input socket at the bottom of the meter. Switch the meter ON by pressing the POWER/ESC button. Prepare a conductivity Standard Solution (1413µS/cm, for example). Hold the probe by its handle and immerse the sensing electrode completely in the standard solution.
The upper display indicates the TDS value (measured in ppm). Rinse the probe with distilled water after each use. Long press the FUNCTION button to return to the conductivity mode. Solution Temperature The lower display indicates the temperature of the measured solution. Refer to the Setup Mode section for instructions on selecting F temperature units and for adjusting the automatic temperature compensation.
Temperature Units Selection The lower display shows C or F. Use the ▲ button to select the desired unit of measure, press ENTER to save. The lower display will show OFF YES or NO. Continue below: Auto Power OFF Enable/Disable The display will show OFF YES or OFF NO.
Specifications General Specifications Circuit Custom one-chip LSI microprocessor circuit Display Dual function LCD Dimensions: 1.7 x 1.1 in. (44 x 29 mm) Measurements Conductivity, TDS, temperature Data Hold Freezes displayed reading MIN-MAX Record and recall lowest and highest readings Auto power off Meter switches off after 10 minutes (can be disabled) Sample rate 1 reading per second (approx.)
Two-year Warranty Teledyne FLIR warrants this Extech brand instrument to be free of defects in parts and workmanship for two years from date of shipment. To view the full warranty text please visit: https://www.flir.com/support-center/warranty/instruments/extech-product-warranty/ Calibration and Repair Services Teledyne FLIR offers calibration and repair services for the Extech brand products we sell. We offer NIST traceable calibration for most of our products.