Texmate DI-45U TIGER CUB User Manual

Texmate DI-45U TIGER CUB User Manual

Smart digital meter

Advertisement

Quick Links

DESCRIPTION
The DI-45U TIGER CUB meter is the most effective solu-
tion to a wide range of measurement and control applications.
With front panel push button commands, the user can setup the
DI-45U TIGER CUB to display a wide variety of input signals,
including DC volts and current, AC volts and current (average
and Trms), thermocouple, 2, 3, or 4 wire 100 ohm platinum
RTDs, four or six wire Strain Gauge signals (load cells and pres-
sure transducers), 4-20mA process signals and frequency. The
DI-45U TIGER CUB also has dual input modules for DC Volts
and DC milliamps.
Input signals may be displayed as voltage or current, or
scaled to read directly in engineering units (eg. PSI, GPM, FPS,
RPM, lbs., degrees, etc.). Scaling in ˚F or ˚C is programmable
from the front panel. With user selectable 3 or 16 readings per
second, the DI-45U TIGER CUB provides fast control response,
true peak readings and a non-isolated analog output signal that
tracks the input signal.
An isolated 5 or 10VDC excitation is available to power pres-
sure transducers and load cells, eliminating the need for an exter-
nal power supply. 24VDC excitation is available on certain input
signal conditioning modules to power process transmitters.
Options include from one to four high isolation relays to im-
plement alarm setpoints programmed by the user. These outputs
can be set to operate above or below the setpoint and in a
latched or non-latching mode. Time delays and hysteresis of the
output are digitally settable. Optional non-isolated analog outputs
of 0-10VDC or 0-20mA (4-20mA) are available to drive chart re-
corders or remote displays. The non-isolated analog output is
scalable from the front panel.
The third simultaneous output option available is logic level
TTL non-isolated serial output, which enables the DI-45U TIGER
CUB meter to communicate with PLC's or personal computers
using the external Texmate Logic To RS232C Isolated Converter.
Baud rates are software selectable from 200 to 19200.
MeterBoss™ software is available to facilitate meter setup.
• DI-45U base unit, standard software, 5 digit red LED display, 100 to 120VAC or 200 to 240VAC
power supply, and plug-in screw terminal blocks (Requires input module. See below). ............................................................DI-45U
• Display Option:
High efficiency green LED display ....................................................................................................................................DI-GREEN5
• Metal sheath option for polycarbonate case ..............................................................................................................OP-MTL 96x48
• NEMA 4 lens cover ......................................................................................................................................................OP-N4X/96x48
• Input Options: See page 13.
• Output Options: See page 12.
3/31/00 DI-45U Data Sheet (DI2)
DI-45U TIGER CUB
SMART DIGITAL METER
An Economical and Powerful Intelligent Panel Meter
for Monitoring, Measurement and Control Applications

SPECIFICATIONS

Input Configuration: ........25 single and dual input
Output Options:
Display Range: ................ -19999 to +32766 or -19990
Input Impedance: ............1 Meg Ohm for DCV (Depend-
Internal Resolution: ........16 bit
Common Mode Rejection:....> 35dB at 50Hz
Conversion Mode: ..........Dual slope; bipolar
Accuracy: ........................±(0.05% of reading + 1 digit)
Conversions per sec: ......Typically 3, programmable to 16
Temp. Coefficient:............< 80ppm/˚C
Display: ............................7 segment, 0.56" high LEDs,
Polarity: ............................Auto
Decimal Point:..................User programmable
Annunciators: ..................Front panel; one annunciator
Overload Indication: ......Indication " ----- "
Case Dimensions: ..........1/8 DIN (96x48x117mm) not
Panel Cutout: ..................45x92mm
Case Material: ................Metal sheathed ruggedized plastic
Power Requirements: ....100 to 120VAC or 200 to 240VAC

ORDERING INFORMATION

Texmate, Inc. Tel. (760) 598-9899
modules
Relay ........................Up to two 10A form C and/or
two 5A form A
Analog ......................Non-Isolated 0-20mA (4-20mA)
or 0-10VDC, scalable by user
Communication..........Logic level TTL non-isolated
to +99990 (dummy zero
selected)
ing on input module selected)
5 digit, red or green
per alarm. Annunciators may
also be programmed to
display trend.
including connectors
Order Part No.
Page 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DI-45U TIGER CUB and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Texmate DI-45U TIGER CUB

  • Page 1: Specifications

    Output Options: With front panel push button commands, the user can setup the DI-45U TIGER CUB to display a wide variety of input signals, Relay ......Up to two 10A form C and/or including DC volts and current, AC volts and current (average...
  • Page 2: Front Panel Functions

    This is in addition to our standard DI-45U TIGER CUB meter package and allows the user to display and modify data direct- If an X appears through a digit, it ly on the PC screen.
  • Page 3: Software Logic Tree

    Any new display value Output/Averaging (Refer to pg. 4 Code entered prior to pressing Definitions) Prog. will not be stored. Code 4 Conversion Rate/Input-Linearization/Input Selection (Refer to pg. 4 Code Definitions) 3/31/00 DI-45U Data Sheet (DI2) Texmate, Inc. Tel. (760) 598-9899 Page 3...
  • Page 4 5 Display Linearized Input Signal Dual Channel V1 + V2 RTD Pt 100 3 wire 6 Program Linearizaiton Points Dual Channel 7 Remote display via serial (V1 x V2)/10000 communications Page 4 Texmate, Inc. Tel. (760) 598-9899 3/31/00 DI-45U Data Sheet (DI2)
  • Page 5 DI-45U Displays toggle of [hiL2] and [XXXXX] (5000 is the default value) STEP E Using the , adjust display to the desired value for hiL2 setpoint. Press “Prog”. DI-45U Displays Operational Display 3/31/00 DI-45U Data Sheet (DI2) Texmate, Inc. Tel. (760) 598-9899 Page 5...
  • Page 6 Time Delay.) Press Prog. button to move through Code 3 Press Prog. button to move through Code 4 and press Prog. again to return to operational display. Page 6 Texmate, Inc. Tel. (760) 598-9899 3/31/00 DI-45U Data Sheet (DI2)
  • Page 7 STEP G Adjust Code 3 Digit 2 to a [x1x]. Press “Prog” button to move to Code 4, then press Prog. button once again to return to operational display. 3/31/00 DI-45U Data Sheet (DI2) Texmate, Inc. Tel. (760) 598-9899 Page 7...
  • Page 8: Point Linearization

    10 POINT LINEARIZATION Prior to the actual programming, Texmate recommends that the Step 9 Using the , adjust display to the user define the 10 linearization points.To program the DI-45U to desired reading when 30% of the input is linearize a signal, the procedure is as follows: applied to the DI-45U.
  • Page 9 MeterBOSS, your computer must be equipped with multiple Display 4, 9, or 16 instruments per screen as bargraphs serial ports, or an external port multiplexer such as Texmate's PortMUX. (Call for more information.) Set multiple alarms for each meter which will appear as flags on...
  • Page 10: Functional Diagram

    MODULE SELECTED) Pin 14 & 15 - AC Power Input. These pins are the power input pins of the meter. The standard DI-45U TIGER CUB is precon- figured for 100/120VAC for meters sold in North America. To reconfigure for 200/240VAC operation, remove the meter from...
  • Page 11: Component Layout

    Thermocouple, RTD, Strain gage, Pressure, Pin 13 - +18VDC @ 30mA Power. For Analog Output Circuit. Resistance, Frequency, and others, some with dual inputs. Contact factory for more information. 3/31/00 DI-45U Data Sheet (DI2) Texmate, Inc. Tel. (760) 598-9899 Page 11...
  • Page 12 Display Board. Pin 21 - Ground for Serial communication Serial Communication Output Available Logic level TTL Non-isolated Serial Output. To communicate with PC’s, use the external Texmate logic to RS232 Isolated Converter. Optional Output...
  • Page 13 NOTE: Use of these application circuits is entirely at the risk and responsibility of the user and any user modifica- tion of the meter may at the discretion of Texmate, void the warranty. (See rear page for user's responsibility and warranty details.) SINGLE INPUT SIGNAL CONDITIONING MODULES See individual module data sheets for more information.
  • Page 14 < Increase Decrease > F01: RPM-20 V AC, 5-30 V DC, 3 selectable ranges; 0-2 kHz, 0-20 kHz and 0-200 kHz PT1: Prototype bare board for customer design and build FREQUENCY/RPM Page 14 Texmate, Inc. Tel. (760) 598-9899 3/31/00 DI-45U Data Sheet (DI2)
  • Page 15 R01: Resistance (2-,3-,4- wire) 200 Ω/2K Ω/20K Ω T01: Thermocouple Universal, J, K, R, S or T. 32 pt. Linearization. NOTE: For any other TC, Contact TEXMATE T02: Pt. 100Ω RTD–(2-,3-,4- wire) 0 to 200.0˚C (32.0 to S01: Strain Gage with 5/10 V DC Excitation, 392.0˚F) -200 to 800˚C (-328.0 to +1472.0˚F).
  • Page 16: Case Dimensions

    However, since we have mal use and service for a period of one year from date of shipment. Texmate’s obligations under this no control over the use of our products once they are shipped, NO WARRANTY WHETHER OF MER-...

Table of Contents