RSA5BUP (Options 65, 66, and 6566)
xx
RSA5UP (Option 5566)
RSA5100A and RSA5100B Series Real-Time Signal Analyzers
Digital I and Q Outputs and Zero Span Output Upgrades
ZZZ
Instructions
www.tektronix.com
*P075106100*
075-1061-00

Advertisement

Table of Contents
loading

Summary of Contents for Tektronix RSA5BUP

  • Page 1 RSA5BUP (Options 65, 66, and 6566) RSA5UP (Option 5566) RSA5100A and RSA5100B Series Real-Time Signal Analyzers Digital I and Q Outputs and Zero Span Output Upgrades Instructions www.tektronix.com *P075106100* 075-1061-00...
  • Page 2 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 3: Service Safety Summary

    Service safety summary Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary located in the Service Manual before performing any service procedures. Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
  • Page 4: Kit Description

    Kit description Kit description This kit describes the installation of the following Options: RSA5100B Series. Option 65: Digital I and Q Outputs Option 66: Zero Span Output Option 6566: Digital I and Q Outputs and Zero Span Output RSA5100A Series. Option 5566: Digital I and Q Outputs and Zero Span Output Products RSA5103B.
  • Page 5: Kit Parts List

    (or options) available is controlled with the option key provided in the kit. Quantity Part number Description 075-1061-XX MANUAL, TECH INSTALLATION, RSA5BUP OPTIONS 65, 66, 6566 171-2191-XX MANUAL,TECH; SVCPT-UPG; TEKTRONIX SUPPLEMENTAL INFORMATION SHEET FOR THE PEOPLES REPUBLIC OF CHINA; CHINA ROHS...
  • Page 6: Installation Instructions

    Installation instructions Installation instructions This section contains all procedures needed to install the required components. Minimum tool and equipment list The following tools are required to for installation of this kit. All tools are standard tools that are readily available. Item Name Description...
  • Page 7 Installation instructions 3. Remove the two T15 Torx-head screws that secure the plastic carrying handle to the side of the instrument. (It is not necessary to remove the black metal handles.) (See Figure 1.) 4. Remove four T15 Torx-head screws along each side that secure the top and bottom covers to the instrument, and two T20 Torx-head screws near the front edge of the top cover (next to the folding handles).
  • Page 8 Installation instructions Figure 1: Remove covers RSA5100A/B Series Digital I/Q and Zero Span Upgrades...
  • Page 9: Attach Cables To Rear-Panel Cover Insert

    Installation instructions Attach cables to rear-panel cover insert 1. Install the cables to the new rear-panel cover provided in this kit. (See Figure 2.) Install the two IQ Output ribbon cables onto the rear panel cover using four T8 Torx-head screws provided in the kit. Tighten the screws using a torque driver set to 5.5 in-lb.
  • Page 10: Install Rear Panel Cover Insert Into Instrument

    Installation instructions Install rear panel cover insert into instrument 1. Remove the blank, rear-panel cover from the chassis (attached with eight T15 Torx-head screws). Discard the blank panel. 2. Install the new rear-panel cover (with attached cabling) on the rear panel chassis using eight T15 Torx-head screws.
  • Page 11: Install Option Board

    Installation instructions Install option board 1. Slide the option board (Digital I and Q / Zero Span) into Slot 3 (counting from the rear of the instrument). (See Figure 4.) CAUTION. Be careful not to bend the pins on the Interface Board when installing the option board.
  • Page 12: Reinstall Top Shield And Cosmetic Covers

    Installation instructions Figure 5: Connecting the output cables Reinstall top shield and cosmetic covers 1. Place the instrument on its bottom feet. 2. Place the top shield on top of the instrument and reinstall the eighteen T15 Torx-head screws. 3. Place the instrument on its rear feet, so the front panel is facing up and the top is toward you.
  • Page 13: Install Option Key

    Installation instructions 8. Align the four screw holes on each side in the top and bottom covers with the holes in the chassis, and install eight T15 Torx-head screws, four on each side. Torque these screws to 8.0 in/lb. 9. Position the plastic carrying handle and its bracket on the right side of the instrument, and install the two T15 Torx-head screws that secure it in place.
  • Page 14: Attach Labels

    Installation instructions 4. Enter the option key provided by Tektronix, and follow the on-screen instructions to install the option. 5. Power off the instrument, then power back on. 6. In the Help menu, select About Tektronix Real-Time Analyzer. 7. Verify the new option is listed.
  • Page 15: Verify Instrument Software Version

    The Export Control label applies only to the RSA5100B Series instruments when installing Option 65 or Option 6566. Verify instrument software version Tektronix recommends to update to the latest available application software for your instrument. 1. Select Help > About to check the software version.