Measurement Guide PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable & Antenna Analyzer Option MW82119B Anritsu Company Part Number: 10580-00402 490 Jarvis Drive Revision: E Morgan Hill, CA 95037-2809 Published: December 2017 Copyright 2014, 2017 Anritsu Company http://www.anritsu.com...
Some or all of the following five symbols may or may not be used on all Anritsu equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this manual.
Page 4
These hazardous compounds present a risk of injury or loss due to exposure. Anritsu Company recommends removing the battery for long-term storage of the instrument and storing the battery in a leak-proof, plastic container. Follow the environmental storage requirements specified in the product technical data sheet.
Here, you can find sales, customer service, and support contact information for your country or region, provide online feedback, complete a “Talk to Anritsu” form to have your questions answered, or obtain other services offered by Anritsu. Updated product information can be found on the Anritsu web site: http://www.anritsu.com/...
(in dBm) and the measured PIM power (in dBm) is the relative power, which is then expressed in units of dBc. The use of dBc units is not applicable to Noise Floor measurements, because the Note transmitters are not On during this measurement. PN: 10580-00402 Rev. E MW82119B MG...
1-4 Why Test for PIM? Why Test for PIM? Anritsu has developed the PIM Master to verify and troubleshoot Passive Intermodulation (PIM). The PIM Master generates two high-power tones, usually in the transmit band of interest. It displays and measures the third-order, fifth-order, or seventh-order intermodulation products returning from the DUT to the PIM Master.
Page 12
• Leave protective caps on RF connectors until you are ready to attach the mating cable. • Avoid loose metal objects within the half-power beam widths of base station antennas, cable trays, vent pipes, air conditioning units, metal flashing, guy wires, and so forth. PN: 10580-00402 Rev. E MW82119B MG...
Figure 1-2 on page 1-7 for a work flow chart of the recommended site testing procedure. Typically, a PIM versus Time test is used to certify PIM performance. Anritsu recommends that you perform a Noise Floor measurement before you begin PIM testing. Refer to Chapter 3, “PIM versus Time”...
A PIM Master equipped with Option 331 is able to perform all site PIM tests and Sweep tests using a single instrument. Start Sweep Site Master Repair Test Noise PIM Master Floor Static Dynamic Save PIM vs. PIM vs. Results Time Time Repair Repair Figure 1-2. Recommended Testing Procedure Work Flow MW82119B MG PN: 10580-00402 Rev. E...
Diplexer Dual band antenna Locations where 900 MHz test tones are injected Locations where IM2 can be measured PIM Locations that could generate harmful IM2 PIM on the antenna surface PIM beyond the antenna PN: 10580-00402 Rev. E MW82119B MG...
Page 17
Option 902) connects the IM2 In port to the 1900 MHz feed line. While transmitting 900 MHz signals, the PIM Master measures the magnitude of the IM2 products arriving at the 1900 MHz BTS. MW82119B MG PN: 10580-00402 Rev. E...
Page 18
PIM Testing for IM2 Products PIM test port (900 MHz test tone Tx) IM2 In test port (Rx only) Locations where 900 MHz test tones are injected Locations where IM2 can be measured 1-10 PN: 10580-00402 Rev. E MW82119B MG...
Sweep Tools can generate a standard report showing plots of the measured results. LST can also generate a special PIM report to display PIM versus TIME or Swept PIM results in a tabular format. Figure 1-5. Example Report Formats MW82119B MG PN: 10580-00402 Rev. E 1-11...
Page 20
1-7 Test Reports Chapter 1 — General Information 1-12 PN: 10580-00402 Rev. E MW82119B MG...
Chapter 2 — Typical Analyzer Setup The Anritsu PIM Master is capable of producing 80 Watts of RF power in the cellular communications bands. Users must take precautions to minimize exposure to these RF fields: Always terminate the PIM output port of the test equipment into a load, a loaded line, or a line that will radiate or absorb the energy before beginning a PIM test.
To turn on the MW82119B PIM Master, press the On/Off button on the front panel. The PIM Master takes approximately forty seconds to complete power-up and to load the application software.
(3rd, 5th, or 7th). 5. Press the Setup main menu key, 6. Set the Output Power level and Test Duration. 7. Press the Setup main menu key to set the power level and test duration. MW82119B MG PN: 10580-00402 Rev. E...
Page 24
11. Visually inspect all RF connectors on the test instrument, test lead, PIM standard, and Low PIM termination. Clean connections each day prior to first use. Save your instrument settings in a setup file to allow easy recall of frequently used Note settings. PN: 10580-00402 Rev. E MW82119B MG...
Verify again After you have made corrections, press this button to continue the calibration. b. Bypass verification Press this button to ignore the warning and continue the calibration. c. Exit calibration Press the button to abort the current calibration. MW82119B MG PN: 10580-00402 Rev. E...
Page 26
Low PIM Termination. If the Low PIM Termination has measured PIM worse than –88 dBm, then the WARNING: High PIM detected dialog box (as shown in Figure 2-1) is displayed. Figure 2-1. Bad Load Detection Warning PN: 10580-00402 Rev. E MW82119B MG...
4. The peak PIM value for the PIM Standard that you are testing should be within ± 3 dB of the numbers shown in Table 2-1 on page 2-8. If the measured PIM is outside of the ± 3 dB limit, contact Anritsu customer service. http://www.anritsu.com/contact.asp MW82119B MG PN: 10580-00402 Rev. E...
Page 28
–116 dBc Typical values are shown. PIM Standards can vary ± 3 dB due to manufacturing Note variation. Record the starting value of your PIM Standard, and use that value for test equipment verification. PN: 10580-00402 Rev. E MW82119B MG...
Verifying the PIM Test Cable When testing, cables are connected and disconnected many times. In order to save wear on the PIM test port cables and adapters, Anritsu recommends removing the O-rings. This allows getting a sufficiently tight connection without unnecessary stress on the connectors.
The IM products of interest are in the same frequency range that is used by mobile user equipment to communicate with the base station. It is therefore possible for nearby mobile equipment to generate signals that are high enough to interfere with your PIM measurement. 2-10 PN: 10580-00402 Rev. E MW82119B MG...
2-4 Calibrating the PIM Analyzer Distance-to-PIM Distance-to-PIM (DTP) is similar to Distance-to-Fault (DTF), which Anritsu introduced in the Site Master for identifying the location of impedance mismatches in a feed line. DTP quickly and accurately identifies the location of PIM faults inside the feed system as well as beyond the antenna.
4. Enter the file name using the touch keyboard (and Quick Name Matrix, if desired) and press Enter (touch screen or hard key). After the Save Location and File Type are changed, these settings will remain in Note effect until manually changed again. 2-12 PN: 10580-00402 Rev. E MW82119B MG...
Calibration Reminder Reset Swept Calibration Trace Overlay Set Default Limit Save Back Other Calibrations Measurement User-Defined Preset Start OSL Preset Return Loss Calibration Save Setup Recall Back Setup Figure 2-2. PIM Analyzer Menu Map MW82119B MG PN: 10580-00402 Rev. E 2-13...
Then enter a name and save the current measurement to file. The file type defaults to measurement, and the appropriate extension is added, based on the current measurement mode. Figure 2-3. Measurements Menu 2-14 PN: 10580-00402 Rev. E MW82119B MG...
Figure 2-4. Test Submenu Key Highlighted in Red Also, during PIM vs. Time and Swept PIM measurements, RF ON is displayed in red in the PIM Summary Table (See Figure 3-3 on page 3-6). MW82119B MG PN: 10580-00402 Rev. E 2-15...
Set Default Limit: Pressing this submenu key sets the active limit line to the default limit line value, depending upon which limit is active. The inactive limit line is not altered. Figure 2-5. Limit Menu 2-16 PN: 10580-00402 Rev. E MW82119B MG...
Back: Press the Back submenu key to return to the previous menu. Figure 2-6. Calibration Menu During calibration, RF is On, but the RF ON message is not displayed. The Note CALIBRATION IN PROCESS... message is displayed in red letters on the measurement screen. MW82119B MG PN: 10580-00402 Rev. E 2-17...
Back: Press the Back submenu key to return to the Back on page 2-17. Figure 2-7. Other Calibrations Menu User Defined Dialog Box Key Sequence: Shift > Cal (2) > Other Calibrations > User-Defined Figure 2-8. User-Defined Calibration Dialog Box 2-18 PN: 10580-00402 Rev. E MW82119B MG...
Chapter 3 — PIM versus Time The Anritsu PIM Master is capable of producing 80 Watts of RF power in the cellular communications bands. Users must take precautions to minimize exposure to these RF fields: Always terminate the PIM output port of the test equipment into a load, a loaded line, or a line that will radiate or absorb the energy before beginning a PIM test.
Noise variation is distributed over a larger number of samples, which results in reduced peak-to-peak variation for low signal level measurements. Trace Mode = FAST Trace Mode = LOW NOISE Figure 3-1. Fast and Low Noise Trace Modes PN: 10580-00402 Rev. E MW82119B MG...
2. Connect the DUT to the test port connector of the PIM Master. Confirm that connections are secure. High power RF signals are emitted from the PIM test port and can cause bodily injury. Anritsu recommends using a torque Warning wrench for this connection.
Page 42
17. Press Test Duration to set the number of seconds from 1 s to 1200 s. 18. Press the Trace Mode submenu key (if desired) to toggle between Fast and Low Noise modes. For additional information, refer to Section 3-2 “Trace Mode” on page 3-2. PN: 10580-00402 Rev. E MW82119B MG...
Page 43
The peak PIM values are shown in red (and remain in red) because the measurement exceeded the limit at some time during the measurement. MW82119B MG PN: 10580-00402 Rev. E...
RF is being transmitted. The Test submenu key, in which the word Measure is underlined while the measurement is in progress, is also highlighted in red while RF is present. PN: 10580-00402 Rev. E MW82119B MG...
Note the PIM bar graph at the left edge of the sweep window. In this graph, the instantaneous PIM level is displayed in units of dBc by the red bar on the vertical axis. No limit was set for this particular measurement. MW82119B MG PN: 10580-00402 Rev. E...
Page 46
MHz submenu key or the Enter submenu key, touch the Continue button, which returns to the same previous menu as when pressing the Back submenu key. Figure 3-5. PIM Parameters Dialog Box PN: 10580-00402 Rev. E MW82119B MG...
The Save dialog box opens. 4. Type a name for the measurement to be saved and press Enter. For additional information about file handling, refer to File Management in your PIM Master User Guide. MW82119B MG PN: 10580-00402 Rev. E...
Antennas. The only problem is that this work was done with FACTORY testing in mind. The IEC specification does not currently address how to apply dynamic stress in the field. Operators, therefore, need to define their preferred test methods. The MW82119B PIM Master has the PIM versus Time measurement mode, which supports dynamic testing and keeps track of the worst case value for the duration of the test.
In this example, the measurement remains below the limit for the duration of the test. The Pass Indicator displays PASS in green letters. Note that the measured PIM and Peak PIM values are displayed in white numerals. Figure 3-7. PASS Indication in PIM vs. Time Measurement MW82119B MG PN: 10580-00402 Rev. E 3-11...
Low Noise Peak Distance-to- Band Select Units Marker High Valley Swept Option PIM Aid PIM Aid Option All Markers Band Select Band Select PCS/AWS Save Measurement Figure 3-8. PIM vs. Time Menu Map 3-12 PN: 10580-00402 Rev. E MW82119B MG...
Figure 3-9. Frequency Menu The left-most main menu key is labeled Freq in the PIM vs. Time and the Note Swept PIM measurement settings. It is labeled Distance in the Distance-to-PIM measurement setting. MW82119B MG PN: 10580-00402 Rev. E 3-13...
Units dBm dBc: Toggles the measurement display units between dBm and dBc. The amplitude for upper and lower limits automatically adjusts to the equivalent level for the selected units. Figure 3-10. Amplitude Menu 3-14 PN: 10580-00402 Rev. E MW82119B MG...
PIM Aid: Press this submenu key to display the PIM Parameters dialog box (see Figure 3-5 on page 3-8). This dialog box displays and allows editing of the current carrier frequencies and carrier power. Figure 3-11. Setup Menu for PIM vs. Time MW82119B MG PN: 10580-00402 Rev. E 3-15...
Marker Valley All Markers Off: Turns off all displayed markers. All Markers Figure 3-12. Marker Menu Select Marker Figure 3-13. Marker Selection Box 3-16 PN: 10580-00402 Rev. E MW82119B MG...
PIM measurement. The Noise Floor Time View measurement displays the signal level that is received by the MW82119B PIM Master at the IM product frequency as a function of time with TX and TX turned Off. The PIM Master is in PIM vs. Time analyzer mode during a Noise Floor - Time View measurement, but only the PIM Master receiver is being used.
14. Select either the Time View or Spectrum View measurement option. 15. Press the Test submenu key to begin the test. While the test is in progress, the Test submenu key is highlighted in red. See Figure 2-4 on page 2-15. PN: 10580-00402 Rev. E MW82119B MG...
PIM testing. Note that units of dBc are not applicable to Noise Floor measurements. Noise Floor Measurement with No External Interference Figure 4-1. Typical Noise Floor (No External Interference, No Limit Line) MW82119B MG PN: 10580-00402 Rev. E...
–97 dBm PIM test limit. This external signal level would be high enough to prevent an accurate PIM test. The test frequencies must be changed to find an IM product frequency without external interference (see Figure 4-3). Figure 4-2. Noise Floor Measurement with External Interference PN: 10580-00402 Rev. E MW82119B MG...
Figure 4-3. Typical Noise Floor Spectrum View Measurement Saving a Noise Floor Measurement To save a Noise Floor measurement, press Shift and File (7), then press the Save or the Save Measurement submenu key. MW82119B MG PN: 10580-00402 Rev. E...
Save Measurement: Press this submenu key to save a Noise Floor measurement. Back Back: Press this submenu key to return to the “Measurements Menu” on page 2-14. Figure 4-4. Noise Floor Menu Measurements Menu Refer to “Measurements Menu” on page 2-14. PN: 10580-00402 Rev. E MW82119B MG...
This capability eliminates the guesswork involved in isolating PIM sources and speeds site repairs. The Anritsu PIM Master is capable of producing 80 Watts of RF power in the cellular communications bands. Users must take precautions to minimize...
Start and Stop distances can also be set in the Distance menu. Cable Loss, Note Propagation Velocity, and Cable specifications can also be set in the DTP Setup menu. Output Power can also be set in the Setup menu. PN: 10580-00402 Rev. E MW82119B MG...
Distance-to-PIM measurement of a PIM standard and a low PIM termination right after calibration. Distance: 0 ft to 100 ft Figure 5-1. DTP Measurement of PIM Standard and Termination after Calibration MW82119B MG PN: 10580-00402 Rev. E...
5. From the DTP Aid menu, press the Units submenu key to display the Units submenu. Press the Meters or Feet submenu key to select the desired units to be used in the measurement. The menu display returns to the DTP Aid menu. PN: 10580-00402 Rev. E MW82119B MG...
Page 65
13. Press the Window submenu key to open the Windowing menu. Select the desired windowing format by pressing one of the four submenu keys: Rectangular, Nominal Side Lobe, Low Side Lobe, or Minimum Side Lobe. Refer to Appendix B, “Windowing”. MW82119B MG PN: 10580-00402 Rev. E...
Page 66
If you need to terminate the measurement before the allotted Test Duration time is complete, then you can press the Test submenu key to turn off the measurement. For additional actions, refer to Section 5-5 “Making a DTP Measurement” on page 5-7. PN: 10580-00402 Rev. E MW82119B MG...
After each measurement, Marker 1 is always placed at the peak value of the DTF measurement, and Delta Marker 1 is always placed at the peak of the DTP measurement. MW82119B MG PN: 10580-00402 Rev. E...
If the distance value is negative, then the unknown PIM source is before the antenna, inside the feed system. Refer to Chapter Figure 5-4. DTP/DTP Trace Overlay Example – PIM Source 10 Ft from End of Cable PN: 10580-00402 Rev. E MW82119B MG...
PIM source is 18.65 ft beyond the end of the cable, indicating that the site PIM is likely beyond the antenna. Figure 5-5. DTP/DTF Trace Overlay Example – PIM Source Beyond End of Cable MW82119B MG PN: 10580-00402 Rev. E...
PIM faults in the sector. The red bars seen here are the DTP enhanced resolution feature described in Section 5-7 on page 5-12. Figure 5-6. DTP/DTF Trace Overlay Example 5-10 PN: 10580-00402 Rev. E MW82119B MG...
Chapter 5 — Distance-to-PIM (DTP) 5-6 Trace Overlay Overlay Using Line Sweep Tools Software After you leave the site, Anritsu Line Sweep Tools (LST) also provides a method to overlay multiple DTP results. Figure 5-7 shows an overlay of two DTP measurements from the main receive path and the diversity receive path of a sector.
PIM sources can be resolved. Table 5-1 shows the calculated resolution for several of the MW82119B frequency bands, assuming a propagation velocity factor of 0.88. As can be seen in the table, some frequency bands have inherently better resolution than others.
Page 73
Chapter 5 — Distance-to-PIM (DTP) 5-7 Enhanced Resolution Figure 5-8. Example of Impulse Bars with Enhanced Resolution MW82119B MG PN: 10580-00402 Rev. E 5-13...
DTP Aid Marker Trace Option Trace Overlay Low Resolution Units Marker Table Band Select Large PCS/AWS Save All Markers DTP Aid More Option Measurement Band Select Resolution Figure 5-9. DTP Analyzer Menu Map 1 5-14 PN: 10580-00402 Rev. E MW82119B MG...
Page 75
Low Side Lobe Calibration Down Bottom Back Window Minimum Side Lobe List Display Favorites Reference Select/Deselect Back Back Favorite Amplitude –100 dBm Clear All Favorites Set Default Reference Figure 5-10. DTP Analyzer Menu Map 2 MW82119B MG PN: 10580-00402 Rev. E 5-15...
Units: Press this submenu key to toggle the units setting between meters and feet. The selection is underlined. Band Select More: Press this submenu key to open the DTP Setup menu. PCS/AWS Option Band Select Figure 5-11. Distance Menu 5-16 PN: 10580-00402 Rev. E MW82119B MG...
Figure 5-2 on page 5-4. The left-most main menu key is labeled Freq in the PIM vs. Time and the Note Swept PIM measurement settings. It is labeled Distance in the Distance-to-PIM measurement setting. MW82119B MG PN: 10580-00402 Rev. E 5-17...
Setup). Units Menu Units Meters Meters: Press this submenu key to set the units to meters. Feet: Press this submenu key to set the units to feet. Feet Figure 5-12. DTP Aid Menu 5-18 PN: 10580-00402 Rev. E MW82119B MG...
Figure 5-13. DTP Setup Menu Settings for Cable Loss, Propagation Velocity, and the Cable list box can also be initiated by highlighting their values in the DTP Parameters window. See Figure 5-2 on page 5-4. MW82119B MG PN: 10580-00402 Rev. E 5-19...
See Figure 5-15 on page 5-21. Favorites Figure 5-14. Cable List Menu You can add cables to the list, remove cables, and modify cable specifications by using Line Sweep Tools. 5-20 PN: 10580-00402 Rev. E MW82119B MG...
Chapter 5 — Distance-to-PIM (DTP) Favorites (in Cable List) Key Sequence: Distance > More > Cable > Display Press Display, if necessary, until Favorites is underlined in the submenu key. Figure 5-17. Cable List Box Showing Favorites 5-22 PN: 10580-00402 Rev. E MW82119B MG...
Minimum Side Lobe Minimum Side Lobe: Press this submenu key to select Minimum Side Lobe windowing. Back Back: Press this submenu key to return to the DTP Setup menu. Figure 5-18. DTP Windowing Menu MW82119B MG PN: 10580-00402 Rev. E 5-23...
Toggles the measurement display units between dBm and dBc. The amplitude for upper and lower limits automatically adjusts to the equivalent level for the selected units. Figure 5-19. DTP Amplitude Menu 5-24 PN: 10580-00402 Rev. E MW82119B MG...
DTP Aid menu. Figure 5-20. DTP Setup Menu The Output Power setting can also be initiated by highlighting its value in the DTP Parameters window. Refer to Figure 5-2 on page 5-4. MW82119B MG PN: 10580-00402 Rev. E 5-25...
Backspace: Press this submenu key (after keying in a test duration with the number keypad) to delete the last digit. Backspace Figure 5-21. DTP Test Duration (Units) Menu Measurements Menu Refer to “Measurements Menu” on page 2-14. 5-26 PN: 10580-00402 Rev. E MW82119B MG...
Large. The Large marker display shows only the x-axis and y-axis values of the active marker. Large All Markers Off: Turns off all displayed markers. All Markers Figure 5-22. Marker Menu Select Marker Figure 5-23. Marker Selection Box MW82119B MG PN: 10580-00402 Rev. E 5-27...
Enter key to use those cable specifications for your measurement. Figure 5-25. DTP Cable Menu DTP Parameters window Cable edit box selected via Arrow keys Figure 5-26. Cable List Submenu Key 5-28 PN: 10580-00402 Rev. E MW82119B MG...
After a valid DTP trace is stored in overlay memory, this submenu key only toggles the view Off and Measurement Figure 5-27. Trace Menu – Distance-to-PIM MW82119B MG PN: 10580-00402 Rev. E 5-29...
Up/Down arrow keys. Set Default Reference: Pressing this submenu key sets the reference limit line to the default limit line value. Set Default Reference Figure 5-28. Reference (Limit) Menu 5-30 PN: 10580-00402 Rev. E MW82119B MG...
Page 91
Chapter 5 — Distance-to-PIM (DTP) 5-22 Reference (Limit) Menu MW82119B MG PN: 10580-00402 Rev. E 5-31...
Page 92
5-22 Reference (Limit) Menu Chapter 5 — Distance-to-PIM (DTP) 5-32 PN: 10580-00402 Rev. E MW82119B MG...
DTP/DTP Trace Overlay. When the trace has been copied to overlay, the trace color immediately changes to the color that has been set in the Select Color choice box (see Figure 6-5 on page 6-6). MW82119B MG PN: 10580-00402 Rev. E...
Page 94
Trace Overlay can be toggled On and Off from the Trace menu or from the Note Measurements menu. Refer to Section 5-21 “Trace Submenu Key” on page 5-29 Section 6-4 “Trace Menu, Distance-to-PIM” on page 6-5. PN: 10580-00402 Rev. E MW82119B MG...
Due to the higher resolution available with DTF traces, each RF connector in the line is clearly visible. In this case, the highest PIM source is clearly coming from the middle RF connector. Figure 6-2. High resolution DTF Measurement Overlaid with DTP Measurement MW82119B MG PN: 10580-00402 Rev. E...
DTP/DTF overlay function. Both measurements identify the RF connector locations, but the reduced resolution here (due to the filter inside the PIM test instrument) is evident over this short test length. Figure 6-3. Automatic DTP/DTF Overlay PN: 10580-00402 Rev. E MW82119B MG...
DTF trace by expanding or contracting its vertical scale. 60.0 dB Figure 6-4. Trace Menu – Distance-to-PIM Moving the top or bottom of the DTF trace allows you to visually compare the distances between key peaks of DTP versus DTF. MW82119B MG PN: 10580-00402 Rev. E...
When On is Max Hold underlined, the trace level will only be updated if the new signal level is greater than the last signal level for each frequency. Figure 6-6. Trace Menu – Spectrum View PN: 10580-00402 Rev. E MW82119B MG...
Chapter 7 — Swept PIM The Anritsu PIM Master is capable of producing 80 Watts of RF power in the cellular communications bands. Users must take precautions to minimize exposure to these RF fields: Always terminate the PIM output port of the test equipment into a load, a loaded line, or a line that will radiate or absorb the energy before beginning a PIM test.
Confirm that connections are secure. High power RF signals are emitted from the Warning PIM test port and can cause bodily injury. Anritsu recommends using a torque wrench for this connection. This device (DUT) may be the main feeder cable from the tower or a simple jumper cable.
RF signal from the PIM Master is On) by using the keypad or the Arrow keys. Then press Enter. The maximum time is 1200 seconds. Calibration If calibration is needed, refer to section “Calibrating the PIM Analyzer” on page 2-5. MW82119B MG PN: 10580-00402 Rev. E...
Test submenu key so that Off is underlined. 3. Save the current measurement by pressing the Save Measurement submenu key. The Save dialog box opens. 4. Type a name for the measurement to be saved and press Enter. PN: 10580-00402 Rev. E MW82119B MG...
IM signal into one of the nulls, then the trace would have indicated that this site was very good, when it is actually bad. Using Swept PIM is a method to verify this condition. MW82119B MG PN: 10580-00402 Rev. E...
Band Select Distance-to- Marker High Valley Swept Swap Swept PIM Reset All Marker Trace To Default Band Select Option PCS/AWS Save All Markers Back Measurement Band Select Option Figure 7-2. Swept PIM Analyzer Menu Map PN: 10580-00402 Rev. E MW82119B MG...
Step Size x.xxx kHz Intermod Order: Select the 3rd, 5th or 7th order of intermodulation. The Anritsu PIM Master automatically sets the center frequency based on the Intermod Order new intermod setting. This submenu key is not displayed when using Option 902.
Reset All To Default: Press this submenu key to reset the carrier Start and Reset All Stop frequencies to full range. To Default Back: Press this submenu key to return to the “Frequency (Freq) Menu”. Back Figure 7-4. Swept PIM Aid Menu PN: 10580-00402 Rev. E MW82119B MG...
Range for the test is 1 second to 1200 seconds. Use the Arrow keys or the number keypad. If you use the number keypad, then the Units menu of the Test Duration menu is displayed. Figure 7-7. Setup Menu for Swept PIM 7-10 PN: 10580-00402 Rev. E MW82119B MG...
All Markers Off: Turns off all displayed markers. All Markers Figure 7-8. Setup Menu for Swept PIM Select Marker Figure 7-9. Marker Selection Box Measurements Menu Refer to “Measurements Menu” on page 2-14. MW82119B MG PN: 10580-00402 Rev. E 7-11...
Page 110
7-9 Marker Menu Chapter 7 — Swept PIM 7-12 PN: 10580-00402 Rev. E MW82119B MG...
These four measurements were taken on a line with two PIM sources that were spaced 100 feet apart. Rectangular windowing provides the highest resolution, or ability to distinguish closely spaced PIM sources. MW82119B MG PN: 10580-00402 Rev. E...
Page 114
Distance To PIM Figure B-1. Rectangular Windowing Total distance is 200 feet. The PIM sources are separated by 100 feet. This view of Rectangular Windowing shows the maximum side lobe display and the greatest waveform resolution. PN: 10580-00402 Rev. E MW82119B MG...
Total distance is 200 feet. The PIM sources are separated by 100 feet. This view of Nominal Side Lobe Windowing shows less side lobe resolution than Rectangular Windowing and more side lobe resolution than Low Side Lobe Windowing. This level of windowing displays intermediate resolution. MW82119B MG PN: 10580-00402 Rev. E...
This view of Low Side Lobe Windowing shows less side lobe resolution than Nominal Side Lobe Windowing and more side lobe resolution than Minimum Side Lobe Windowing. This level of windowing displays intermediate resolution. PN: 10580-00402 Rev. E MW82119B MG...
Total distance is 200 feet. The PIM sources are separated by 100 feet. This view of Minimum Side Lobe Windowing shows less side lobe resolution than Low Side Lobe Windowing and displays the lowest side lobe and waveform resolution. MW82119B MG PN: 10580-00402 Rev. E...
Page 118
B-2 Distance-to-PIM (DTP) Windowing Examples Appendix B — Windowing PN: 10580-00402 Rev. E MW82119B MG...
Page 124
W to W Index-6 PN: 10580-00402 Rev. E MW82119B MG...