Servicing explosion-protected versions

After acknowledging using the
8 seconds remain until a new error mess-
age can appear.
The short time interval is long enough to
start the transfer of new data from the mem-
ory pen.
The instrument is automatically reset on re-
moving the memory pen.
With communication using a PC or note-
book, the instrument is automatically reset
after the data is transferred.
Other errors
These errors must be acknowledged by
pressing the
key to allow the instrument
to continue working.
28
EB 9527-1 EN
key,
9 Servicing explosion-protected
versions
In the event that a component of the Media
6 on which the explosion protection is
based must be serviced, the instrument must
not be put back into operation again until
an expert has inspected the device accord-
ing to explosion protection requirements,
has issued a certificate stating this, or given
the device a mark of conformity.
Inspection by an expert does not have to be
carried out, if the manufacturer performs a
routine check test on the device prior to tak-
ing it into operation again, and the success
of the routine check test is documented by
attaching a mark of conformity to the de-
vice.
Explosion-protected components may only
be replaced by original checked compo-
nents from the manufacturer.