c Mix or check that the disinfection chemicals have the concentration recommended by the manufacturer
(for example, a chemical strip test).
d Check that the temperature of the disinfectant is within the manufacturer's recommended limits.
5 Perform a high-level disinfection of the TRANSDUCER using the following procedure:
a Immerse the transducer in a high-level disinfectant solution.
WARNING:
Do not soak the transducer longer than recommended by the chemical
Caution:
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a non-
Connector
Do not submerge
b Ensure that the connector and most of the cable remain out of the fluid. You may submerge 5 cm (2
inches) of the cable proximal to the transducer scanhead.
6
Rinse the transducer 3 separate times using the following procedure:
112
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer's rinse instructions to remove
chemical residue.
manufacturer.
recommended disinfecting solution or incorrect solution strength can damage or
discolor the transducer and void the warranty.
Cable
Determining the required cleaning and disinfection level
Maximum
5cm
(2 inches)
Scanhead