Hardware-related errors are usually indicated by an error message. Use the following table to identify
hardware-related errors and their remedies.
Table 9-2: System error messages
Error message
The signal path has
stopped because the
transducer is not
connected.
The temperature sensor is
not working, either in the
transducer or the system.
The system temperature is
too high.
The transducer
temperature is too high.
Software-related errors typically display an error message on the screen.
If the system displays a software error
1 Stop the scan, and return to the Home screen.
2 From the Home screen, tap Scan or Patient to see if the system has recovered.
If the error persists or another error appears
1 Turn off the system and install a fully charged battery. Be sure to save any patient information before
turning off the system.
2 Press the Power button to turn on the system.
3 If the system powers on normally, it has recovered from the fault, and you may use your system.
If the error persists
1 Write down the error message.
2 Press and hold the Power button.
3 When prompted, tap Power off, and then OK.
Troubleshooting
Remedy
Reattach the transducer.
Note that it is important to end an exam before disconnecting
the transducer.
Disconnect and reattach the transducer, and try again.
Turn off the system, and allow it to cool for 10 minutes before
turning it on again.
Turn off the system, and disconnect the transducer. Allow the
transducer to cool for 10 minutes, and try again.