Operation
34
Menu item
Measurement /
Temp. comp. (TC) /
Linear coeff.
Measurement /
Temp. comp. (TC) /
Reference temp.
Measurement /TDS
factor
Measurement /
Stability control
Measurement /
Temperature unit
Reset
Possible
Description
setting
0.000 ...
Coefficient of the linear
3.000 %/K
temperature compensation.
This setting is only available
when the linear temperature
compensation is set.
20 °C
Reference temperature
25 °C
This setting is only available
for the measured
parameters,
0.40 ... 1.00
Factor for TDS value
On /Off
Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)
°C
Temperature unit,
°F
degrees Celsius or degrees
Fahrenheit.
All temperatures are
displayed with the selected
unit.
-
Resets all sensor settings to
the delivery condition (see
section 4.6.1).