Cond 3210
ba75794e04
04/2012
Menu item
Measurement /
Measuring cell /
Type
Measurement /
Measuring cell /
Cell const. man
Measurement /
Measuring cell /
Cell const. LR325/01
Measurement /
Temp. comp. (TC) /
Method
Possible
Description
setting
Measuring cell used
Cal
Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution.
Calibration ranges:
0.450 to 0.500 cm
0.800 to 0.880 cm
The currently valid cell
constant is displayed in the
status line.
LR325/01
LR 325/01 measuring cell,
nominal cell constant
0.100 cm
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm
LR325/001
LR 325/001 measuring cell,
nominal cell constant
0.010 cm
The cell constant is
permanently set.
man
Any measuring cells with
freely adjustable cell
constants in the range from
0.250 to 25.000 cm
0.250 to
Display and setting option of
-1
25.000 cm
the cell constant of any
measuring cells (man).
0.090 to
Display and setting option of
-1
0.110 cm
the cell constant of the
LR 325/01 measuring cell.
nLF
Procedure for temperature
lin
compensation (see section
Off
4.4.3).
This setting is only available
for the measured
parameters,
Operation
-1
and
-1
-1
.
-1
.
-1
.
-1
.
ӕ
ρ
und
.
33