Maintenance Schedule - DENTSPLY CEREC MC X Cleaning And Maintenance

Table of Contents

Advertisement

Maintenance Schedule

Before every wet machining process
• Ensure that the water and Dentatec solution completely covers
the filter(s) in the water tank
Weekly
• Change water in water tank and add 75ml Dentatec
Monthly
• Clean water filter monthly or as needed depending on
usage (change filter every 3-4 cleanings)
• Check and clean water jets to ensure that water is striking
the tools during wet machining
• Clean manual block clamp inside of the block axle
• Clean motor clamping cones for instruments without chuck
Yearly
• Annual maintenance carried out by a Service Technician
Dry milling zirconia only:
Every 10 dry milled zirconia restorations
• Vacuum chamber and run wet cleaning program
Every 120 dry milled zirconia restorations
• Change filter bag (change HEPA filter every 3-4 filter bags)