Gynnie® OB/GYN Stretcher
1061
Operations Manual
2017/08 A.4
1061-109-001 REV A
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker Gynnie 1061

  • Page 1 Gynnie® OB/GYN Stretcher 1061 Operations Manual 2017/08 A.4 1061-109-001 REV A www.stryker.com...
  • Page 2 sample text...
  • Page 3 General warning Caution Warning; crushing of hands No pushing Do not store the oxygen bottle Catalogue number Serial number Batch code For US Patents see www.stryker.com/patents CE mark EC REP Manufacturer Date of manufacture Safe working load Lubricate www.stryker.com 1061-109-001 REV A...
  • Page 4 sample text...
  • Page 5: Table Of Contents

    Cleaning..............................22 Cleaning the product.......................... 22 Cleaning the mattress ........................22 Remove iodine..........................23 Special instructions ........................... 23 Disinfecting............................24 Disinfecting the product ........................24 Disinfecting the mattress ........................24 Preventive maintenance ........................... 26 Warranty ............................... 27 www.stryker.com 1061-109-001 REV A...
  • Page 6 Table of Contents Warranty exclusion and damage limitations .................... 27 To obtain parts and service ......................... 27 Return authorization........................... 27 Damaged product..........................27 International warranty clause ....................... 27 1061-109-001 REV A www.stryker.com...
  • Page 7: Warning/Caution/Note Definition

    Note: Provides special information to make maintenance easier or important instructions clearer. www.stryker.com 1061-109-001 REV A...
  • Page 8: Summary Of Safety Precautions

    • Do not allow fluid to pool on the mattress. Fluids can cause corrosion of components and may cause the safety and performance of this product to become unpredictable. 1061-109-001 REV A www.stryker.com...
  • Page 9 Always make sure that there is no interference between the waste bin holder and objects in the base storage tray to avoid damage to the product or the objects. • Always make sure that the waste bin holder is secure before you attach a disposable bag, basin, or instrument tray. www.stryker.com 1061-109-001 REV A...
  • Page 10: Introduction

    Introduction This manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating or maintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenance of this product.
  • Page 11: Specifications

    66 x 191.5 cm Siderails 13 x 39 in. 34 x 99 cm 8 in. 20 cm Caster diameter 68 in. 172.7 cm Break away from wall point Stryker reserves the right to change specifications without notice. www.stryker.com 1061-109-001 REV A...
  • Page 12: Product Illustration

    Foot section IV pole Foot section release handle Foot support Pump pedal Uni-lower pedal Fowler backrest Contact information Contact Stryker Customer Service or Technical Support at: 1-800-327-0770. Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 1061-109-001 REV A www.stryker.com...
  • Page 13: Serial Number Location

    Contact information (Continued) To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. Serial number location...
  • Page 14: Setup

    Raise and lower the manual gatch (foot section). 10. Extend and retract each footrest and make sure that you can adjust the footrest position and lock the footrest. 11. Raise and lower each push handle and make sure that the push handle locks. 1061-109-001 REV A www.stryker.com...
  • Page 15: Operation

    To lower the foot end of the litter, press down on the side of the uni-lower pedal (B) closest to the foot end of the product. Note: The base may be equipped with optional variable descent controls. With variable descent controls, the farther you press down on the pedal, the faster the litter will lower. www.stryker.com 1061-109-001 REV A...
  • Page 16: Positioning The Product In Trendelenburg

    Note: Raise the litter to the highest height for a greater Trendelenburg angle. To lower the foot end of the product, push down on the side of the uni-lower pedal (B) closest to the foot end (Figure 2 on page 12). 1061-109-001 REV A www.stryker.com...
  • Page 17: Transporting A Patient With The Retractable Fifth Wheel

    Push down on the handles to lock them into position. Reverse steps to stow the handles. Note: Only use the push handles as push or pull devices unless otherwise specified to avoid product damage. Figure 3: Positioning the head end push handles www.stryker.com 1061-109-001 REV A...
  • Page 18: Raising Or Lowering The Siderails

    To raise the Fowler backrest, squeeze the red Fowler backrest release handle (A). Pull the Fowler backrest up to the desired position. To lower the Fowler backrest, squeeze the red Fowler backrest release handle (A). Push the Fowler backrest down to the desired position. Figure 4: Fowler backrest 1061-109-001 REV A www.stryker.com...
  • Page 19: Positioning The Foot Supports

    • Do not use the foot supports as a push/pull device. Damage to the product may occur. • Do not place large items in the base storage tray. Damage to the foot support may occur. www.stryker.com 1061-109-001 REV A...
  • Page 20: Positioning The Foot Section

    Slide the waste bin holder (A) out from under the litter frame (Figure 6 on page 17). Place the waste basin in the holder. Secure the top of the basin on the “wings” (B) of the waste bin holder. 1061-109-001 REV A www.stryker.com...
  • Page 21 Positioning the waste bin holder (optional) (Continued) To store the waste bin holder, remove and dispose of the waste basin. Slide the waste bin holder under the litter frame. Figure 6: Position the waste bin holder www.stryker.com 1061-109-001 REV A...
  • Page 22: Accessories

    Accessories These accessories may be available for use with your product. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770. Name Number Calf rest, free standing 1061-040-036 Calf support mounting brackets 1061-130-000 IV pole, removable 0390-025-000 1020-130-000...
  • Page 23: Attaching The Upright Oxygen Bottle Holder

    Insert the support bar into any of the IV sockets. Insert the cotter pin through the hole in the support bar to secure the bottle holder to the product. Note: Do not use the upright oxygen bottle holder as a push/pull device. www.stryker.com 1061-109-001 REV A...
  • Page 24: Extending Or Stowing The Serving Tray Holder/Footboard

    Reverse steps to secure the X-ray cassette in the mounting brackets (C). To remove the tray from the Fowler, lift the bottom of the tray out of the mounting brackets (D). Figure 8: Accessing the Fowler X-ray cassette holder 1061-109-001 REV A www.stryker.com...
  • Page 25: Locating The Patient Restraint Strap Tie-Ins

    There are six patient restraint strap tie-in locations on the litter assembly for attaching patient restraint straps (Figure 9 on page 21). Figure 9: Restraint strap tie-in locations www.stryker.com 1061-109-001 REV A...
  • Page 26: Cleaning

    Use a clean, soft cloth to wipe down the entire mattress with a mild soap and water solution to remove foreign material. Wipe down the mattress with a clean, dry cloth to remove any excess liquid or cleaning agents. 1061-109-001 REV A www.stryker.com...
  • Page 27: Remove Iodine

    Use standard household cleansers or vinyl cleansers and a soft bristle brush on troublesome spots or stains. Pre- soak dried-on soil. Laundering Laundering is not recommended. Laundering may substantially decrease the useful life of the mattress. www.stryker.com 1061-109-001 REV A...
  • Page 28: Disinfecting

    Phenolic disinfectant (active ingredient - o-phenylphenol) • Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available chlorine (40 mL of a 5.25% bleach solution per 4000 mL water)) 1061-109-001 REV A www.stryker.com...
  • Page 29 • Frequent or prolonged exposure to higher concentrations of disinfectant solutions may prematurely age the cover fabric. • The use of accelerated hydrogen peroxides or quaternaries that contain glycol ethers may damage the cover. www.stryker.com 1061-109-001 REV A...
  • Page 30: Preventive Maintenance

    Remove product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
  • Page 31: Warranty

    Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by Stryker, product or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper use or any alteration or repair by others in such a manner as in Stryker’s judgment affects the product materially and adversely, shall void...