ABB FCM2000 Operating Instruction page 36

Coriolis mass flowmeter for simultaneous measurement of mass, density and temperature
Table of Contents

Advertisement

Submenu
Contact input
Switch contacts
Totalizer reset
Contact output
No function
Submenu
TAG Number
Label
123ABCDE
Descriptor
123456789ABCDE
Date
07.07.2001
Instrument no.
1234567
Submenu
Communication
Data link
HART
Instr. address
0
36
FCM2000
Parameter
Contact output
Max Alarm _ _
* only visible for Technician
Value Range/
Entry Type
no function
Sets current and pulse outputs to
Ext. zero return
0% flowrate
Resets all mass and volume totalizers
Totalizer reset
(see Page 57 Paragraph 6.6.16)
Contact is open
This menu is used to define the function of
no function
the contact output.
Contact open
F/R-Signal_
when no forward flow
⇒ Contact closed
F/R-Signal /
when no forward flow
⇒ Contact open
General alarm _
when no general alarm
⇒ Contact closed
General alarm /
when no general alarm
⇒ Contact open
MAX/MIN Alarm _
when no MAX/MIN Alarm
⇒ Contact closed
MAX/MIN Alarm /
when no MAX/MIN Alarm
⇒ Contact open
MIN Alarm _
when no MIN-Alarm
⇒ Contact closed
MIN Alarm /
when no MIN-Alarm
⇒ Contact open
MAX Alarm _
when no MAX-Alarm
⇒ Contact closed
MAX Alarm /
when no MAX-Alarm
8 ASCII
ASCII
HART TAG
(see Page 58 Paragraph 6.6.17)
16 ASCII
HART Descriptor
1.1.1900 -
HART Date
31.12.2155
0-999999
display
HART
Selection of the communication mode
off
(see Page 59 Paragraph 6.6.18)
(Technician) *
0 - 15
Integer
In this menu the instrument address is de-
fined. For HART-Communication ad-
(Technician) *
dresses must be between 0 and 15.
! Information:
If HART is selected as the communication
mode and an instrument address other
than 0 is entered, the converter will oper-
ate in the so called Multidrop-Mode. In
this mode current output 1 (HART-current
output) is set to a constant value of 4 mA.
6 Converter
Comments
D184B111U02