ABB FCM2000 Operating Instruction page 31

Coriolis mass flowmeter for simultaneous measurement of mass, density and temperature
Table of Contents

Advertisement

Damping
5.0 s
Low flow cutoff
2.1 %
System zero adj.
-0.0111%
System zero adj.
automatic?
System zero adj.
manual
Submenu
Alarm
Min-Alarm Qm
0.00%
Max-Alarm Qm
100.00%
Min Alarm Density
0.5 kg/l
Max Alarm Density
3.5 kg/l
Min Alarm Temp.
-50.0 °C
Max Alarm Temp.
100.00 °C
D184B111U02
FCM2000
Parameter
System zero adj.
slow?
System zero adj.
fast
System zero adj.
0.011%
* only visible for Technician
Value Range/
Entry Type
1-100 s
float
Damping for current /pulse outputs;
Response time for a step change to reach
(Technician) *
99 % of its final value
0 - 10 %
float
(based on Qmax)
(Technician) *
(see Page 45 Paragraph 6.6.8)
Using „DATA" or „STEP" select „manual"
or „automatic"
(see Page 45 Paragraph 6.6.9)
function Automatic system zero adjustment;
slow,
fast
float
Technician menu
Manual entry of a
system zero value
0-105 %
float
Lower mass flowrate limit must be less
(Technician) *
than the upper mass flowrate limit.
The entry is based on QmMax.
0-105 %
float
Upper mass flowrate limit must be greater
(Technician) *
than the lower mass flowrate limit.
The entry is based on QmMax.
0.5-3.5 kg/l
float
Lower density limit must be less than the
upper density limit.
(Technician) *
0.5-3.5 kg/l
float
Upper density limit must be greater than
the lower density limit.
(Technician) *
-50 to 180 °C
float
Lower temperature limit must be less than
(Technician) *
the upper temperature limit.
-50 to 180 °C
float
Upper temperature limit must be greater
(Technician) *
than the lower temperature limit.
6 Converter
Comments
31