ABB FCM2000 Operating Instruction page 34

Coriolis mass flowmeter for simultaneous measurement of mass, density and temperature
Table of Contents

Advertisement

Totalizer reset
Submenu
Pulse output
Output of
Mass
Qvmax pulse
100.00 l
Pulse factor
631
Pulse width
30.000 ms
Submenu
Current output 1
Output
Qm
Qv → l = 100%
120.00 l/min
Density → l = 0%
500.000 g/l
Density → l=100%
1000.00 g/l
Temp. → l = 0%
-10.0 °C
Temp. → = 100%
180 °C
34
FCM2000
Parameter
Overflow ← R
0
reset?
yes -> ENTER
/kg
Value Range/
Entry Type
display
Display overflow counter, reverse flow;
Max. 65535 overflows;
1 overflow = 10,000,000
Only displayed when Operating Mode is
Forward/reverse
function
Reset all mass and volume totalizers
(see Page 50 Paragraph 6.6.13)
Mass
float
The units are selected in the Menu „Unit
Volume
Totalizer"
0.0001 to
1000
Pulses/Unit
Only displayed for „Output of Volume"
Display of the pulses to be outputted for
each selected mass or volume totalizer
unit
Only displayed for „Output of Mass".
float
(see Page 51 Paragraph 6.6.14)
Qm
float
Selection of the value to be indicated ny
the current output.
Qv
Qmax = 20 mA
Density
Temperature
0.1 to 1000000
Only displayed for „Output of Volume".
Volume flowrate for a current output of
100%
0 to 1000000
float
Only displayed for „Output of Density".
Density for a current output of 0%
0.1 to 1000000
float
Only displayed for „Output of Density".
Density for a current output of 100%
-50 to 190°
float
Only displayed for „Output of Temp.".
Temp. for a current output of 0%
-50 to 190°
float
Only displayed for „Output of Temp.".
Temp. for a current output of 100%
6 Converter
Comments
D184B111U02

Advertisement

Table of Contents
loading

Table of Contents