Aquamaster Explorer Input/Output Connections - ABB AquaMaster Explorer Instruction Manual

Electromagnetic flowmeter
Hide thumbs Also See for AquaMaster Explorer:
Table of Contents

Advertisement

Electromagnetic Flowmeter
AquaMaster™ Explorer

3.4.3 AquaMaster Explorer Input/Output Connections

Fig. 3.10 AquaMaster Explorer – Bulgin Connectons (Connector, Remote version)
Fig. 3.11 AquaMaster Explorer – MIL Style Connectons (Connector, Remote version)
Pin
Pin
(Bulgin)
(MIL Style)
1
A
2
B
3
C
4
D
5
E
6
F
7
G
8
Table 3.1 AquaMaster Explorer – Connector Input/Output Connections
IM/AM/E Issue 10
WABC 2020/01
WABC 2010/01
Signal
+V
Scanreader +V
DATA
Scanreader Data
O/P COM
Output Common
O/P2
Reverse Pulses or Direction Indicator
O/P3
Alarm Output
O/P1
Forward Pulses or Forward & Reverse Pulses
0V
Scanreader 0V
NC
Not Used
Pin View
1
7
8
6
5
Pin View
B
C
Function
3 Electrical Installation
2
3
4
A
F
G
E
D
Color
(Output Cable)
Violet
Blue
Yellow
Red
Brown
Orange
Screen
21