Model 4200A-SCS Parameter Analyzer User's Manual
Section 2: Getting started
Figure 8: Interlock connector on the rear panel of the 4200A-SCS
2. Connect the other end of the interlock cable to a compatible test fixture, such as the Keithley
Instruments LR:8028.
For more information, see the "Configuring the safety interlock" topic in the 4200A-SCS Reference
Manual.

SMU connections

The following topics explain how to connect the source-measure units (SMUs) to the device under
test (DUT).
Do not touch test cables or connectors when powering up the 4200A-SCS. Hazardous
voltage may be output momentarily, posing a safety hazard that could result in personal
injury or death.
Do not turn on the 4200A-SCS until you have reviewed the safe power-up procedure in
Powering the 4200A-SCS
(on page 2-16).
Do not connect the DUT to the 4200A-SCS before powering it up, because the hazardous voltage
that may be output momentarily at power-up could damage the DUT.
If your 4200A-SCS includes preamplifiers, all tests should be performed using the preamplifiers, as
the installed SMUs were optimized at the factory to use them.
4200A-SCS-900-01 Rev A. / September 2016
2-11