Advertisement

Quick Links

Protos
3400(X)
®

Additional Functions

Latest Product Information:
www.knick.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Protos 34009(X) and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Knick Protos 34009(X)

  • Page 1: Additional Functions

    Protos 3400(X) ® Additional Functions Latest Product Information: www.knick.de...
  • Page 2 Calimatic® Protos® Sensocheck® Sensoface® ServiceScope® VariPower® are registered trademarks of Knick GmbH & Co. KG, Germany Memosens® is a registered trademark of Endress+Hauser Conducta GmbH & Co. KG, Germany Knick Elektronische Messgeräte GmbH & Co. KG, Germany SMARTMEDIA® is a registered trademark of Toshiba Corp., Japan Knick Elektronische Messgeräte GmbH &...
  • Page 3 Additional Functions for the Protos 3400(X) modular measuring system Additional functions How to order an additional function ................... 5 Information on TAN release and SmartMedia cards Activating an additional function (Option release) ............6 SmartMedia card ........................32 Provided by: TAN SW 3400-001 KI recorder ......................
  • Page 4 Protos 3400(X)
  • Page 5: How To Order An Additional Function

    How to Order an Additional Function Device-specific TAN (transaction number) Additional functions expand the device capabilities. The additional functions are device-specific. When ordering an additional function, you therefore have to specify the serial number of your FRONT module in addition to the respective order number. (The FRONT module contains the Protos system control).
  • Page 6 Activating an Additional Function Select menu: Parameter setting/System control/Release of options Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number (see previous page). Menu Display Activate additional function 0.003mS/cm Menu selection 25.6 °C Open parameter setting.
  • Page 7: Sw 3400-001 Ki Recorder

    SW 3400-001: KI Recorder Select menu: Parameter setting/Module FRONT/KI recorder Additional function SW 3400-001 The KI recorder follows the course of the process and releases a messages in the event of a deviation. It is called from the measuring mode: meas Continuous processes The measured signal is surrounded by a dynamically tracked, process- controlled tolerance band.
  • Page 8 SW 3400-001: KI Recorder Configuring the KI recorder Menu Display °C °C KI recorder Called directly from the measuring mode: Press meas key. For function description, KI recorder °C press “Info” softkey 8.56 29.4 °C Info 07/09/10 11:23 Info on KI recorder “Info”...
  • Page 9 SW 3400-001: KI Recorder Menu Display KI recorder: Batch process Configuring the KI recorder 7.00 pH 25.6 °C • Parameter setting of FRONT module KI recorder (Administrator) • Select KI recorder KI recorder Channel 1 Continuous process • Select the variables to be displayed Channel 2 Batch process as well as the start, end values,...
  • Page 10 SW 3400-001: KI Recorder Display KI recorder: Batch process Recording a reference curve Start “Record” with left softkey. Reference value and envelope are displayed during the recording process. At the end of recording or after “KI stop” the KI recorder °C switches to control mode.
  • Page 11 SW 3400-001: KI Recorder KI recorder: Batch process Display Evaluation Within the “KI recorder” function you can evalu- ate a reference curve directly on the device. To do so, use the arrow keys (up/down). Evaluation channel 1: The reference curves and the most important 9.00 pH 8.35 pH Start...
  • Page 12 Protos 3400(X)
  • Page 13 SW 3400-002: Specifiable Buffer Set Select menu: Parameter setting/System control/Buffer table Individual buffer set (with 3 buffer solutions) for pH measurement Buffer table You can enter an individual buffer set (with 3 buffer solutions). To do so, you enter the nominal buffer values for the correct temperature (range 0 ... 95 °C / 5°C steps).
  • Page 14: Sw 3400-004 Servicescope

    SW 3400-004: ServiceScope (pH) Select menu: Diagnostics/Module PH/ServiceScope ServiceScope ServiceScope monitors the pH input signal and checks whether it remains in the control range. An error message is generated if the input signal exceeds the failure limit. Menu Display ServiceScope ServiceScope 7.00 pH 20.2 °C...
  • Page 15 SW 3400-005: Tolerance Adjustment (pH) Select menu: Diagnostics/Module PH/Tolerance band recorder Tolerance band recorder (For adjusting the tolerance band, see next page.) During manual calibration this function checks the zero and slope values and automatically performs an adjustment when the tolerance band is exceeded. The parameters are stored in the tolerance band recorder (Diagnostics menu).
  • Page 16 SW 3400-005: Tolerance Adjustment (pH) Adjustment of tolerance band. Select menu: Parameter setting/Module PH/Cal preset values/Cal tolerance band Menu Display Parameter setting Menu selection 7.00 pH 17.1 °C From the measuring mode: Menu selection Press menu key to select menu. Select parameter setting using arrow keys, press enter to confirm.
  • Page 17: Sw 3400-006 Current Characteristic Definable

    SW 3400-006: Current Characteristic Definable Select menu: Parameter setting/Module BASE (OUT)/Output current/ Characteristic Menu Display Specifying a current characteristic Table characteristic Assign output current to measured 7.03 pH value in 1-mA steps. 25.6 °C Output current I1 (Administrator) Select menu: Parameter setting/ Variable Module BASE (OUT)/Output current Curve...
  • Page 18 Current Curve Adjustments – Original for Copy Parameter setting/Module BASE (OUT)/Output current/ Characteristic "Table", own adjustments Value at measured variable 00 mA ...._________ 01 mA ...._________ 02 mA ...._________ 03 mA ...._________ 04 mA ...._________ 05 mA ....
  • Page 19: Sw 3400-008 Tc Ultrapure Water (Cond)

    SW 3400-008: TC Ultrapure Water (COND) Select menu: Parameter setting/Module COND/TC process medium/ Ultrapure water Menu Display TC process medium Ultrapure water with traces of 0.20mS/cm 22.3 °C impurity TC process medium (Administrator) • Parameter setting EN 27888 Natural waters (0 ... 35°C) Ultrapure water: traces of impurity •...
  • Page 20: Sw 3400-009 Concentration Determination

    SW 3400-009: Concentration Determination Select menu: Parameter setting/System control/Module COND/Concentration Default settings and selection range Concentration determination The substance concentration in percent by weight (% by wt) is determined for H , HNO , HCl, NaOH, NaCl from the measured conductivity and temperature values.
  • Page 21 Parameter Setting: Concentration Curves (Additional function SW 3400-009) Default settings and selection range Note: HOLD mode active Parameter Default Selection / Range On, Off Concentration (0-30 %), H (32-84 %), H (92-99 %), • Medium (“Yes” selected) (0-30%) (0-30 %), HNO (35-96 %), HCl (0-18 %), HCl (22-39 %), NaOH (0-14 %), NaOH (18-50 %),...
  • Page 22 Nitric acid HNO c [% by wt] Hydrochloric acid HCl c [% by wt] Protos 3400(X)
  • Page 23 Sodium hydroxide solution NaOH c [% by wt] Table salt solution NaCl c [% by wt]...
  • Page 24 SW 3400-009: Concentration Table Select menu: Parameter setting/System control/Concentration table Specifying a concentration solution for conductivity measurement Concentration table To specify the customer-specific solution, 5 concentration values A-E are entered in a matrix together with 5 temperature values 1-5. To do so, first enter the 5 temperature values, then enter the respective conductivity values for each concentration A-E.
  • Page 25 The table is built up as 5 x 5 matrix: Conc. A Conc. B Conc. C Conc. D Conc. E Temp 1 Temp 2 Temp 3 Temp 4 Temp 5 Conditions for the table: • The temperature values must be rising (Temp 1 is the lowest, Temp 5 the highest temperature).
  • Page 26 SW 3400-011: Dissolved Oxygen Measurement ... Application-specific additional function for breweries This additional function simplifies parameter setting since all steps not required for dissolved oxygen measurement in carbonated beverages are omitted. It simultaneously acts on all installed OXY modules (module software version 2.2 and higher). Function principle: The following processes are automated by the additional function, i.e.
  • Page 27 Parameter Default Selection / Range Input filter Off, Weak, Medium, Strong, On (000 sec... 999 sec) Weak • Pulse suppression Sensor data • Measure in Liquids A Standard, B Trace sensor (with guard) • Sensor type A Standard C Trace sensor (without guard), Others NTC 30 kohms, NTC 22 kohms •...
  • Page 28: Sw 3400-012 Isfet For Ph 3400(X)-035 Ism Module

    SW 3400-012: ISFET for PH 3400(X)-035 ISM Module Connection for ISM sensor (e.g. Mettler-Toledo InPro 3300) PH 3400(X)-035 Jumper Note: Each time a new sensor is connected, you must perform an ISFET zero adjustment to adjust the operating point. After that, you should perform one of the following calibration methods: •...
  • Page 29 Menu Display Display Parameter settings: New Parameter setting / Module 7.00pH 20.1 °C PH 3400-035 / Sensor data Module PH 3400-035 (Administrator) Sensor data are preset depending on Input filter Sensor data the sensor type. Cal preset values Gray display lines cannot be edited. TC process medium ORP/rH value Delta function...
  • Page 30: Sw 3400-014 2Nd Channel For Fiu 3400-141, Fiu 3400X-140

    SW 3400-014: 2nd Channel for FIU 3400-141 / FIU 3400X-140 Module FIU 3400-141 module Channel1 (Memosens, radio transmission or probe control) The 2nd channel for connecting a Memosens sensor or enabling Channel 2 radio transmission allows you (Memosens or radio transmission, to make use of the full range of activation via TAN) functions of the FIU module –...
  • Page 31: Sw 3400-015 Oxy Measurement For Memosens Modules

    SW 3400-015: OXY Measurement for Memosens Modules MS 3400(X)-160, FIU 3400-141, FIU 3400X-140 SW 3400-016: OXY Trace Measurement for Memosens Modules MS 3400(X)-160, FIU 3400-141, FIU 3400X-140 Memosens module Memosens PH or Memosens COND, FIU modules including: radio transmission or probe control TAN required: OXY: SW 3400-015 OXY traces: SW 3400-016...
  • Page 32: Smartmedia Card

    Inserting the SmartMedia Card To release an additional function via TAN, see Pg 6: Release of options Inserting and replacing the SmartMedia card The SmartMedia card may be inserted or replaced with the power supply switched on. Protect against electrostatic discharge! When closing the device, make sure that the sealing is properly seated and clean.
  • Page 33 SMARTMEDIA Card: Usage Use as memory card in combination with additional functions. Additional functions must be ordered separately (activation via TAN). A SmartMedia card is provided with the following additional functions: • Software update (SW 3400-106, see Pg 45) • Memory card (SW 3400-102 ... 1xx (see Pg 36 and the following) Using commercially available SmartMedia cards Commercially available SmartMedia cards can be used as additional memory card.
  • Page 34 Saving / Loading Device Configuration System control/Memory card/Copy configuration. Saving / loading the complete device configuration System control/Memory card/Copy configuration. “Save“ configuration means that the complete device configuration (except the passcodes) is written on the memory card. “Load” configuration means that the complete device configuration is read from the memory card and pogrammed.
  • Page 35 SMARTMEDIA Card: Formatting a Card Parameter setting/System control/Format card Formatting a card Menu Display (generating a memory card) To format a card 0.003mS/cm 20.4 °C • Insert SmartMedia card (Pg 32) Software update (Administrator) • Open menu selection Updating will change the device characteristics.
  • Page 36: Memory Card

    SMARTMEDIA Card: Memory Card Parameter setting/System control/Memory card Menu Display Using the memory card Using the memory card 0.003mS/cm 20.4 °C • Insert SmartMedia card (Pg 32) Software update (Administrator) • Open menu selection Updating will change the device characteristics. Perform •...
  • Page 37 SW 3400-102: Loadable Parameter Sets Parameter setting/System control/Parameter sets Loading a parameter set from Menu Display SmartMedia card Loading a parameter set from 0.003mS/cm 25.1°C SmartMedia card Parameter sets (Administrator) • Open parameter setting For switchover see "Function control matrix" •...
  • Page 38 SW 3400-102: Loadable Parameter Sets Parameter setting/System control/Parameter sets Saving a parameter set on Menu Display SmartMedia card To save a parameter set on 0.003mS/cm 25.1°C SmartMedia card Parameter sets (Administrator) • Open parameter setting For switchover see “Function control matrix” •...
  • Page 39: Sw 3400-103 Measurement Recorder

    SW 3400-103: Measurement Recorder Select menu: Parameter setting/Module FRONT/Measurement recorder The last 200 data records are stored in the analyzer and are represented graphically – like on a recorder. All further data can be stored on the SmartMedia card and be evaluated by a computer. 2 process variables are represented.
  • Page 40 SW 3400-103: Measurement Recorder Select menu: Parameter setting/Module FRONT/Measurement recorder Additional function SW 3400-103: Measurement recorder The measurement recorder saves all entries in a file. The last 200 entries can be displayed on the Protos. A new file is generated for each day. The date is encoded in the file name.
  • Page 41 SW 3400-103: Measurement Recorder Select menu: Parameter setting/Module FRONT/Measurement recorder Meaning of the entries in the recorder file: Time stamp of recorder entry Left channel of recorder with measured value and unit Right channel of recorder with measured value and unit NAMUR signal* “Maintenance request”...
  • Page 42: Sw 3400-104 Extended Logbook

    SW 3400-104: Extended Logbook Parameter setting/System control/Logbook Additional function SW 3400-104: Extended logbook The extended logbook saves all entries in a file. The last 50 entries can be displayed on the Protos. A new file is generated for each month. The date is encoded in the file name.
  • Page 43: Sw 3400-106 Software Update

    SW 3400-106: Software Update For a software update (additional function SW3400-106), the manufacturer supplies a specially formatted SmartMedia card. The analyzer replaces its own firmware (operating program) by the new version (“Update”). Caution! During a software update the analyzer is not operable! After a software update you should check the configuration.
  • Page 44 Card: Saving Firmware SMARTMEDIA Parameter setting/System control/Software update/Save firmware Saving firmware Menu Display to software-update card To save firmware 0.003mS/cm 20.4 °C • Insert SmartMedia card Software update (Administrator) • Open menu selection Updating will change the device characteristics. Perform •...
  • Page 45 Card: Loading Firmware SMARTMEDIA Parameter setting/System control/Software update/Load firmware Software update Menu Display (“Load firmware”) Software update 0.003mS/cm 25.6 °C • Insert SmartMedia card (Pg 32) Release of options (Administrator) • Open menu selection 102 Parameter sets Inactive 103 Measurement recorder Inactive •...
  • Page 47 AuditTrail and Signature Electronic data recording and backup according to FDA 21 CFR Part 11 Additional function SW 3400-107 Provided by: TAN and AuditTrail card Contents AuditTrail ............................48 Signature ............................49 Activating the AuditTrail ......................50 Signature entry ..........................51 AuditTrail: Parameter setting ....................52 AuditTrail: User management ....................53 Audit Trail: Recording.......................54 AuditTrail log ..........................55...
  • Page 48 Readability from PC is not affected. Note: Only specially coded SmartMedia cards supplied from Knick can be used as AuditTrail cards. As delivered, the AuditTrail card includes a PC program which makes the recorded AuditTrail data readable.
  • Page 49 Signature Function description Signature A signature consists of two entries, the user name and the corresponding pass- code. Each change at the device (calibration/adjustment, parameter setting, maintenance, if required also diagnostics) is recorded in the “AuditTrail Log” and the respective records together with the name of the user who has performed the change.
  • Page 50 Activating the AuditTrail Select menu: Parameter setting/System control/Release of options Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number! Menu Display Activating the additional function 0.003mS/cm Menu selection 25.6°C Open parameter setting. Menu selection From the measuring mode: Press menu key to select menu.
  • Page 51 Entering a Signature (User name, passcode) Entering a signature (user name, passcode) to open a menu To protect the data against unauthorized query or modification, access data must be entered before a main menu from the menu selection is opened. Two-stage security is provided by input of user name and passcode.
  • Page 52 AuditTrail: Parameter Setting Adapt function: Parameter setting/System control/AuditTrail Parameter setting In the System control menu the Administrator can adjust the audit trail system to the requirements of his company. For example, he/she can select the protection against unauthorized query and the encrypting of the data in the audit trail log.
  • Page 53 AuditTrail: User Management Create, edit, erase user data Menu Display AuditTrail: User management Menu selection 7.00 pH 17.1 °C From the measuring mode: Menu selection Press menu key to select menu. Select parameter setting using arrow keys, press enter to confirm. Select: [enter] Signature for first access:...
  • Page 54 AuditTrail Log Function AutoComplete The “AutoComplete” function facilitates the input of the user name. You only have to enter so many letters as the analyzer requires to identify the name. This function can be switched off in the AuditTrail menu because the automatic display of the user name reduces system security, which then might not be sufficient for FDA documentation.
  • Page 55 AuditTrail Log Parameter The audit trail log cannot be queried by the Protos. It can only be evaluated using the suitable PC program. According to FDA 21 CFR Part 11 the user must ensure the transport of the data from the PROTOS to evaluation and further processing or archiving.
  • Page 56 AuditTrail Log Point of measurement Tank UF0815 FRONT 3400-011 0006123 BASE 3400-021 0006458 PHU 3400-110 0007221 OXY 3400-062 0006045 6653 07/06/09 10:43:02 Login: Admin (ADMIN) 6654 07/06/09 10:43:02 Login: > Parameter setting (Administrator) 6655 07/06/09 10:43:02 Login: > [1] Module PHU 3400-110 (Admin.) OK 6656 07/06/09 10:43:02 Login: >...
  • Page 58 Protos 3400...
  • Page 60 Additional Functions Provided by: device-specific TAN To release: Parameter setting / System control / Release of options Passcode: Administrator level 1989 (new: ....) SW 3400-001 KI recorder ................7 SW 3400-002 Specifiable buffer set (pH) ..........13 SW 3400-004 ServiceScope ..............14 SW 3400-005 Tolerance band recorder (pH) ........15 SW 3400-006 Current characteristic definable .........19 SW 3400-008 TC ultrapure water (COND) ..........19...

Table of Contents