Specifications
Accuracy Specifications
Thermocouple Accuracy
Thermocouple Type
J
K
1
Thermocouple measurement specifications include cold-junction compensation error (with sensor between 15° and 35° C),
isothermal accuracy and system noise. The specifications assume that the 0.625 V range is used and that ground-referencing
and open-thermocouple detection are enabled for a floating thermocouple. Specifications improve with ground-referencing
enabled and open-thermocouple detection disabled for a floating thermocouple. The specifications also assume that the
cold-junction sensor is between 15° and 35° C.
© National Instruments Corporation
This appendix lists the specifications of the NI 4350. These
specifications are for 15° to 35° C ambient temperature range for one
year unless otherwise specified. All specifications are relative to
calibration standards and require a 30 minute warm-up period.
Specifications do not include transducer error. Temperature coefficient
is applicable for 0° to 15° and 35° to 55° C. For thermocouples, add the
accessory error in ° C only if the accessory (TC-2190, PSH32-TC6,
CB-27T, TBX-68T) is in the 0° to 15° and 35° to 55° C temperature
range.
1
Error (° C)
15°–35° C, 1 Year
Filter Setting
° C
10 Hz
50 Hz
–100
0.53
0.61
0
0.42
0.49
760
0.42
0.47
–100
0.60
0.72
0
0.45
0.54
1000
0.60
0.69
1372
0.74
0.84
Temperature
Coefficient
60 Hz
° C/° C
0.74
0.02
0.59
0.55
0.89
0.03
0.67
0.81
0.99
A-1
Appendix
A
0°–15° C,
35°–55° C
Accessory
Error
*
° C
0.25