6. Quick reference for making the first
1. Connect the measuring unit ISOMET2114 to the supplied adapter and connect the adapter to
proper mains, according to technical data on the label of the power adapter.
2. The green LED indicator for charging batteries will start flashing continuously.
3. After a certain time, depending on state of the batteries, flashing will stop and the green LED
will light continuously. Steady light of the green LED means, that charging batteries has been
stopped.
4. Disconnect the adapter
5. Connect measurement probe to the measuring unit. At connecting the measurement probe,
follow the instructions in the chapter 3.4 Measurement Probes.
6. Insert the needle probe into the measured material or place the surface probe on the smooth
surface of the measured material.
7. Switch on the measuring unit by pressing the red Power key for approx. 3 seconds. For
further reference see the chapter 4 Operation.
8. After switching the instrument on, the Stand By window will be shown on the display. After
switching the instrument on, the instrument may be connected to the adapter at laboratory
applications or it may be fully powered from rechargeable batteries at outdoor applications.
9. At this stage, continuous measurement of thermal conductivity may be started by pressing the
key F1 on the Stand by page.
10. After running the thermal conductivity measurement, the Stand By page will be changed to
Measurement page and the measurement progress will be indicated by the progress bar on
the bottom of the page.
11. Measurement results are available by pressing the key F4 or by transferring results into the
PC according to instructions in the chapter 5.
12. Transferring results into the PC interrupts the measurement progress.
13. Measurement in progress may be stopped by pressing the key F1 on the Measurement page
and confirming the request for stopping by pressing the key F5. (Further reference in chapter
4.1 Measurement).
14. The switched ON measuring unit is switched OFF by pressing red Power key for approx. 3
seconds. After switching the instrument OFF the display will get dark. The green LED for
indicating battery charger function will light further, if power adapter is connected to the
instrument.
Note:
At certain states of rechargeable internal batteries, if the measuring unit is switched on with
connected power adapter or USB connection with PC, pressing the red Power for 3 seconds may not
switch on the instrument successfully. In this case, the instrument display remains black and the red
LED will be lighting continuously. Such state may be overcome by disconnecting the power adapter
and the USB connection and turning the instrument off by pressing the red Power key for
approximately 10 seconds, while the red LED will switch of. Therefore it is suggested to switch the
instrument on with disconnected power adapter and USB connection.
28
measurement
ISOMET 2114
User's Guide
www.appliedp.com