Table 5. Motor Design Type and Characteristics Summary
Motor Design
NEMA-A
Starting Current
High
Starting Torque
Medium
Breakdown Torque
High
Required from motor nameplate
Rated Power
Rated Speed
Direct on Line
VSD
Rated Voltage
50 Hz motors
60 Hz motors
Rated Current
Rated Frequency
Rated Cos Q (DPF)
Rated Service Factor
Motor Design Type
Default values are available for the nameplate settings and ranges. Use  (UNIT SETUP) to open the
setup screen to configure these values. The selection depends on the default motor frequency.
Softkeys:
Select UNIT SETUP screen
Select TREND SETUP screen
Select DEFAULTS for the
MOTOR SETUP screen
START measurements
NEMA-B
NEMA-C
Medium
Medium
Medium
High
Medium
High
Table 6. Nameplate Settings and Ranges
746 W to 746 kW
1 hp to 1000 hp
300 rpm to 3600 rpm
800 rpm to 3600 rpm
170 V to 1000 V
208 V to 1000 V
1 A to 1500 A
50 Hz or 60 Hz
0.01 to 1.00
1.00 to 1.50
NEMA-A, NEMA-B, NEMA-C, NEMA-D, NEMA-E, IEC-H,
IEC-N, Other
Opens the Motor Analyzer Function Preference screen.
Select metric or imperial units for the mechanical motor
power, torque, and the default motor frequency.
Adjusts the duration of the automatic trend recording. The
default duration is 7 days.
Sets the default values defined in the Function Preferences
screen (see UNIT SETUP). The defaults depend on the
default motor frequency.
Starts the Motor Analyzer function. See Motor Analyzer
Parameters on page 8.
NEMA-D
NEMA-E
IEC-H
Medium
Medium
Medium
Very High
Medium
High
Very High
Medium
High
Range
Motor Analyzer
Motor Setup
IEC-N
Medium
Medium
Medium
Resolution
1 W to 1 kW
1 hp
1 rpm
1 V
0.1 A
- - -
0.01
0.01
- - -
5