The myoMOTION system is intended for use in electromagnetic environment specified below. The customer or the user
of the myoMOTION system should assure that it is used in such an environment.
Immunity Test
Electrostatic
±6 kV contact
discharge (ESD)
± 8 kV air
IEC 64000-4-2
Electrical fast
±2kV for power supply
transient/burst
lines
IEC 61000-4-4
±1kV for input/output
lines
Surge
±1kV differential
mode
IEC 61000-4-5
±2kV common mode
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycle
on power supply
input lines
40 % U
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
For 25 cycles
<5 % U
(>95 % dip in U
For 5 sec
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity
The myoMOTION system is intended for use in electromagnetic environment specified below. The customer or the user
of the myoMOTION system should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
(Charging System)
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
P-6828 Rev A (Oct 2014)
IEC 60601
Compliance level
test level
±6 kV contact
±6 kV air
±2kV for power supply
lines
Not applicable
±1kV differential mode
±2kV common mode
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
Not applicable to
T
)
operation
T
3 A/m
Electromagnetic environment -
Device user should avoid touching subject
and sensor probes while a measurement
is active.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
For battery charging mains power quality
should be that of a typical commercial or
hospital environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Compliance
Electromagnetic environment -
level
Portable and mobile RF communications
equipment should be used no closer to
any part of the myoMOTION system,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
3Vrms
d
1
2 .
d
1
2 .
3V/m
d
2
3 .
myoMOTION Data Logger
guidance
guidance
P
P
80 MHz to 800 MHz
P
800 MHz to 2,5 GHz
29