Section 16: Specifications of the Product

 
Expected Useful Lifetime 
The myoMOTION Data Logger (#682) has a usable life of seven years.
The myoMOTION Data Logger (#682) operates with a rechargeable Lithium Ion battery. The
battery capacity will decline with ongoing use and require replacement after 500+
discharge/charge cycles to preserve the device's rated 8 hours of operating time.
 
Dimensions and Weight 
myoMOTION DataLogger Dimensions
37.6mm (4.13in) L x 65.09mm (2.56in) W x 15.88mm (0.625in) H
myoMOTION Data Logger Weight: 125g (4.4oz)
Performance Characteristics 
Output & Transmission Frequency (Depending on country) 
Up to 2.5 mW (depending on country allowance)
DSSS 2415-2472 MHz on (up to) 18 selectable radio channels
myoMOTION sensor transmission range: 10m (typical)
Signal latency of 140 msec during data collection
MyoMOTION Sensor Data Acquisition System 
X, Y, Z acceleration sampled at 800 Hz
X, Y, Z angular velocity sampled at 800 Hz
X, Y, Z magnetic field sampled at 800 Hz
Sampled sensor values algorithmically combined into four 16-bit quaternion values
Quaternion estimates generated at 100 Hz and 200 Hz (200 Hz only available in the
Research System)
Orientation data (joint angles) are available in both the Clinical and Research
sensors, but only the research system has linear acceleration data available.
Energy Consumption, Condition of Use 
Data Logger is powered by Lithium-Ion battery and charged by a 5V USB host
Environmental Conditions for Storage and Transport 
Ambient Temperature: -40C to +70C
Relative Humidity: 10% to 100%
Atmospheric Pressure: 500hPa to 1060hPa
 
IP (Ingress Protection) Rating 
 
The instrument is not protected against the ingress of water and carries no IPX rating (i.e. is
ordinary equipment). The myoMOTION Data Logger is not waterproof. Care must be taken to
avoid exposure to all liquids.