Exceptional Functions/Situations (Error Messages); Shutdown After Use - Noraxon myoMOTION 682 User Manual

Table of Contents

Advertisement

Noraxon U.S.A., Inc.
Exceptional Functions/Situations (error messages) 
Error Message
Too many sensors
Clinical sensors do not support
acceleration
Clinical sensors only support 100Hz
sample rate
No sensors enabled
The following sensors are not responding:
XXXX
Sensor firmware must be > version X for
sensors
 
Shutdown after Use 
At the end of the day:
Place all myoMOTION sensors inside the sensor charging station
Apply AC wall power to the charging station which disables the sensor radios
Turn the myoMOTION Data Logger off
Connect the myoMOTION Data Logger's USB cable to the computer or USB wall power
supply
 
P-6828 Rev A (Oct 2014)
Meaning
Too many sensors have been enabled or the
sample rate is set too high. The data logger will
support 16 sensors and 2 footswitches at
200Hz sample rate or 8 sensors at 200Hz
sample rate with acceleration enabled
Clinical sensors do not support acceleration for
clinical sensors. Acceleration needs to be
disabled in order for the data logger to work
with clinical sensors.
Clinical sensors do not support 200Hz sample
rate. Sample rate must be changed to 100Hz.
Sensors may have been assigned, but they
have not been enabled. Sensors must be
turned on in order to record data.
The sensors listed are not communicating with
the data logger (XXXX represents the serial
number of the sensors not communicating)
The sensors' firmware needs to be updated to
a certain version to work with the data logger
 
myoMOTION Data Logger