E. Getting Started
1.
Disconnect the charger from the device if connected.
Connect the measurement cable and switch the device On – after a short
2.
logo presentation, the main screen is displayed and the cursor on the tooth
image will start blinking.
(While turning the device On without the measurement cable – the main
screen will be displayed without cursor on the tooth image).
3.
Attach the lip clip to the patient.
Insert the file into the canal – (to ensure precise measurements the file size
4.
should be adjusted to the canal diameter).
5.
Connect the hook to the file.
The cursor on the tooth will stop blinking (accompanied by a double beep signal).
A blinking cursor and no sound signal indicates a faulty connection. Check your
connections, clean your hook connection, moisten the canal, if necessary, and start
again.
No other adjustments are necessary before starting measurement.
Note: Connection test feature is included in ProPex
in order to check the cables:
● Connect metal part of the connection hook
to the lip clip
● "Connection test" icon
in the bottom left corner of the display,
indicating proper connection
● If no icon appears, the connection hook or
the measurement cable should be replaced
®
II
should appear
12 / 120