SPECIALISTS IN THE MEASUREMENT AND CONTROL OF VISCOSITY Boston • Chicago • London • Stuttgart • Guangzhou with o ces in: BROOKFIELD ENGINEERING LABORATORIES, INC. 11 Commerce Boulevard, Middleboro, MA 02346 USA 508-946-6200 or 800-628-8139 (USA e xcluding MA) 508-946-6262 http://www.brook eldengineering.com...
III. OPERATION .......................... 7 III.1 Full Scale Range and Accuracy of Measurement ................17 III.2 Accuracy of Viscosity and Temperature ....................18 III.3 Calibration Verification ..........................20 III.4 Cone Calibration ............................22 III.5 Repeatability ..............................24 III.6 Making Viscosity Measurements ......................25 III.7 Computer Control ............................26 APPENDIX A - Variables in Viscosity Measurement ..................28 APPENDIX B - Communications .........................30 APPENDIX C - Warranty Repair and Service ....................34 This manual intended for use with CAP 2000+ series viscometers which have serial numbers beginning with a prefix of “CPN”. CAP1000 and 2000 Viscometers with a serial number prefix of “CP” require a different manual. Please contact Brookfield or your authorized dealer/representative to obtain this manual. Brookfield Engineering Labs., Inc. Page 2 Manual No. M02-33E209...
UK........DE-8 Germany ......DE-7 6. Operating Instructions Manual ....M02-313 The following optional items may have been included: Part No. 7. Viscosity Standard Fluid for calibration See Table III-5 and III-6 in Section III Brookfield Engineering Labs., Inc. Page 3 Manual No. M02-33E209...
Please check to be sure that you have received all components and that there is no damage. If you are missing any parts, please notify Brookfield or your local dealer immediately. Any ship- ping damage must be reported to the carrier. Save the packing container, if possible, for future use when returning the viscometer to Brookfield or an authorized dealer for service.
Page 5
Temperatures: CAP 2000+L 5°C (or 15°C below ambient, whichever is higher) to 75°C CAP 2000+H 50°C to 235°C All models provide 0.1°C increments Weight: Gross Weight 36 lb 16.3 kg Net Weight 27 lb 12.3 kg Carton Volume 4.9 cu ft 0.15 m...
4) Connect the power cord to the viscometer and to the power supply (source). 5) If using a printer, connect the printer cable to the printer port and printer. Note: Ensure that both the printer and the CAP 2000+ are off when connecting cables. Brookfield Engineering Labs., Inc. Page 7 Manual No. M02-33E209...
This instrument is not intended for use in a potentially hazardous environment. In case of emergency, turn off the instrument and then disconnect the electrical cord from the wall outlet. Brookfield Engineering Labs., Inc. Page Manual No. M02-33E209...
Viscosity is displayed in either P=Poise or cP=Centipoise (CGS system) or Pa•s=Pascal seconds or mPa•s=milliPascal seconds (SI system). If the viscosity measurement is over range, “EEEE” will be displayed. Brookfield recommends a minimum torque reading of 10% when making viscosity measurements. If the torque value is between 0 and 10%, the display will flash to indicate an under range condition.
Solvent Trap (P/N C1K-63) is made of Teflon. Clean with non abrasive cloth and solvent appropriate for sample material that is not aggressive to Teflon. When cleaning, take care not to apply excessive force which may bend the spindle shaft or otherwise damage the instrument. Brookfield Engineering Labs., Inc. Page 0 Manual No. M02-33E209...
For example: if using cone 4 (shear rate multiplier of 3.33), an entered shear rate of 51 1/SEC will be displayed as 50 1/SEC. Brookfield Engineering Labs., Inc. Page Manual No. M02-33E209...
After the correct two (2) digits have been entered, press the ENTER key and the cone will be accepted for viscometer calculations. The screen will display the following message: Brookfield Engineering Labs., Inc. Page 2 Manual No. M02-33E209...
YES and see Section III. Cones supplied at the time of order are calibrated to the viscometer by Brookfield prior to shipment, but should still be verified with a calibration check prior to use for the first time. Cones purchased separately at a later time by the user must also be calibrated prior to making viscosity measurements.
It is recommended to have the spindle in contact with the plate prior to introducing the sample material to ensure that the spindle is also at the temperature of test. Brookfield recommends using the solvent trap at all times to enhance the temperature control of the sample material.
(Section II.1) Pressing and holding the STOP and PRINT keys simultaneously executes the printing of a new heading (Section II.7). Brookfield Engineering Labs., Inc. Page Manual No. M02-33E209...
Note: The viscometer will be operating according to the selected parameters while the parameter display screen is shown. Speed 15 RPM Shear Rate 50 1/sec Run 20 S Hold 10S Spindle 04 Figure II-6 Brookfield Engineering Labs., Inc. Page Manual No. M02-33E209...
III. Full Scale Range and Accuracy of Measurement Full Scale Range (FSR) viscosity is the maximum viscosity that can be measured and occurs when the % torque is 100. Brookfield recommends that viscosity measurements be made be- tween 10 and 100% torque.
% FSR from the measurement. • Consult the table below to determine the accuracy of your measurement • Multiply the accuracy by the FSR viscosity if you need your accuracy stated in Poise. Brookfield Engineering Labs., Inc. Page Manual No. M02-33E209...
Page 19
5°C to 75°C. HIGH TEMP UNITS: • In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point on the viscometer is 50°C to ≤100°C. Brookfield Engineering Labs., Inc. Page 9 Manual No. M02-33E209...
The cones/spindles that were shipped with the viscometer were calibrated to the viscometer by Brookfield prior to shipment. If your viscometer was returned to Brookfield or an authorized Brookfield Dealer, then any cones/spindles that were returned for service will have been cali- brated to the viscometer prior to shipment.
Page 21
6. If the test fails, repeat again to make sure that every step was performed correctly. If the test fails again, perform a cone calibration according to the procedure in Section III.4. Brookfield Engineering Labs., Inc. Page 2 Manual No. M02-33E209...
Viscosity Standard Fluids. This field calibration will accommodate any wear on the tip of the cone which may result from contact with the plate. This also allows the user to purchase a new cone spindle and perform the calibration in your plant Brookfield Engineering Labs., Inc. Page 22 Manual No. M02-33E209...
Page 23
Full Scale Range. Spindles used with the high torque CAP 2000+ and the recommended viscosity standard fluids may be calibrated at any speed. Brookfield Engineering Labs., Inc. Page 23 Manual No. M02-33E209...
However, more weight means more wear on the cone and plate. Additional weights should only be considered when definitely required and removed when not required. Contact Brookfield Engineering Laboratories or your Brookfield Dealer/Distributor/Representa- tive for details on the above information.
Cone calibration is only required when a new cone (i.e., replace- ment for lost/damaged cone) is used, or when calibration check fails. c) When making measurements with low temperature instruments (CAP 2000+L), the solvent trap may not be required (for the containment of solvents and/or prevention of sample “drying”).
The CAP 2000+ Viscometer can be operated remotely under PC control when using the CAP- CALC application software. When advanced sample analysis is required, Brookfield CAPCALC application software can control the CAP 2000+ Viscometer from a PC. CAPCALC for Windows requires Windows Brookfield Engineering Labs., Inc.
Page 27
5°C and 75°C (CAP 2000+L) or 50°C and 200°C (CAP 2000+H) depending on viscometer model. CAPCALC features include on-screen and printed plots of % F.S.R., viscosity or shear stress vs.
If you set a Viscometer at a constant speed, recording P values over time, and find that the P values decrease with time, the material is thixotropic. Brookfield publication, “More Solutions to Sticky Problems” includes a more detailed discus- sion of rheological properties and non-Newtonian behavior. Brookfield Engineering Labs., Inc.
Page 29
3. Test temperature 4. Test speed [or the shear rate] 5. Length of time to record viscosity 6. Sample volume sufficient to cover the face of the cone - Warranty Repair and Service Brookfield Engineering Labs., Inc. Page 29 Manual No. M02-33E209...
When connecting the CAP 2000+ to a computer, use Brookfield Computer Cable (Part No. DVP-80). If you are not using the Brookfield computer cable, jump (connect) pins 4 and 9 (refer to Figure 1) on the CAP 2000 end of the serial cable. The cable connections are:...
Page 31
• yyy : calibration speed. Subject to same limitations as in V command. • If there is an error, bit 7 of the status byte is set. Invalid Com- ???<CR> • Invalid command received mand Table 1 Brookfield Engineering Labs., Inc. Page 3 Manual No. M02-33E209...
Page 32
Speed (RPM) input (hexadecimal). All speeds are three characters padded with leading zeroes. INTERPRETATION OF BYTE STATUS Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Motor Off Motor On Over SFR (>=115.0%) Valid FSR (<115.0%) Value Outside Limits Value Within Limits Calibration Error Brookfield Engineering Labs., Inc. Page 32 Manual No. M02-33E209...
Page 33
1/sec, a sensed temperature of 100.0°C, and a number one cone. Note that Bit 3 of the status byte has been cleared. It is set only in response to the offending command (T9C4) and is cleared as soon as the response is issued. Brookfield Engineering Labs., Inc. Page 33 Manual No. M02-33E209...
Telephone: (508) 946-6200 FAX: (508) 923-5009 www.brookfieldengineering.com For repair or service outside the United States consult Brookfield Engineering Laboratories, Inc. or the dealer from whom you purchased the instrument. For repair or service in the United Kingdom return to: Brookfield Viscometers Limited...
Page 35
PACKAGING INSTRUCTIONS TO RETURN A BROOKFIELD CAP VISCOMETER FOR REPAIR OR CALIBRATION Package the viscometer for shipment as outlined below. Return the viscometer to the attention of the Repair Department (see address below). RUSH SERVICE MUST BE INDICATED ON THIS FORM AND CLEARLY MARKED ON THE OUTSIDE OF YOUR SHIPPING PACKAGE.
Page 36
Please note that we will only contact you if repairs exceed $400.00. ❑ Visa ❑ MasterCard ❑ American Express ❑ Purchase Order No. ___________________________________________ Name on Credit Card: __________________________________ Return Authorization No. Expiration Date _______________________________________ Account No. _________________________________________ Brookfield Engineering Labs., Inc. Page 3 Manual No. M02-33E209...