Brookfield CAP 1000+ Operating Instructions Manual

Brookfield CAP 1000+ Operating Instructions Manual

Viscometer
Hide thumbs Also See for CAP 1000+:

Advertisement

Quick Links

Brookfield Engineering Labs., Inc.
BROOKFIELD CAP1000+ VISCOMETER
MODEL CAP 1000+
Operating Instructions
Manual No.
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346 USA
T
508-946-6200
or 800-628-8139 (USA excluding MA)
EL
F
508-946-6262
I
AX
NTERNET
M02-312
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
http://www.brookfieldengineering.com
Page 1
Manual No. M02-312

Advertisement

Table of Contents
loading

Summary of Contents for Brookfield CAP 1000+

  • Page 1 MODEL CAP 1000+ Operating Instructions Manual No. M02-312 SPECIALISTS IN THE MEASUREMENT AND CONTROL OF VISCOSITY BROOKFIELD ENGINEERING LABORATORIES, INC. 11 Commerce Boulevard, Middleboro, MA 02346 USA 508-946-6200 or 800-628-8139 (USA excluding MA) 508-946-6262 http://www.brookfieldengineering.com NTERNET Brookfield Engineering Labs., Inc.
  • Page 2: Table Of Contents

    “CPN”. CAP1000 and 2000 Viscometers with a serial number prefix of “CP” require a different manual. Please contact Brookfield or your autho- rized dealer/representative to obtain this manual. Brookfield Engineering Labs., Inc.
  • Page 3: Introduction

    3. Spindle Case ......... CAP-106Y 4. Solvent Trap ......... CAP-118 5. Power Cord: 115V ........ DVP-65 220V ........ DVP-66 UK ........DE-8 Germany ......DE-7 6. Operating Instructions Manual ....M/02-312 Brookfield Engineering Labs., Inc. Page 3 Manual No. M02-312...
  • Page 4: Utilities

    Brookfield Engineering or your local Brookfield dealer immediately. Any shipping damage must be reported to the carrier. Save the packing con- tainer, if possible, for future use when returning the viscometer to Brookfield or an authorized dealer for service.
  • Page 5: Installation

    This instrument is not intended for use in a potentially hazardous environment. In case of emergency, turn off the instrument and then disconnect the electrical cord from the wall outlet. Brookfield Engineering Labs., Inc. Page 5 Manual No. M02-312...
  • Page 6: Key Functions

    (SI system). If the viscosity measurement is over range, “EEEE” will be displayed. Brookfield recommends a minimum reading of 10%. If the display value is between 0 and 10%, the unit display will flash to indicate an out of range condition. If the viscometer settles out with a final reading below zero, negative values will be displayed.
  • Page 7: Cleaning

    Note: When cleaning, take care not to apply excessive force which may bend the spindle shafft or otherwise damage the instrument. Brookfield Engineering Labs., Inc. Page 7 Manual No. M02-312...
  • Page 8: Getting Started

    Turn the power ON using the switch located on the rear of the base console. The display will sequentially show BROOKFIELD, then the model of the viscometer and the version number. After about four seconds, the main screen will be displayed (Figure II-1), indicating the temperature of the sample plate.
  • Page 9 Note: 1. CAP Viscometers allow for only one cone at a time of the same cone number to be calibrated to the viscometer. Multiple cones of the same cone number must each be calibrated to the viscometer before operation (refer to Appendix B). Brookfield Engineering Labs., Inc. Page 9 Manual No. M02-312...
  • Page 10: Ii.3 Speed Setting

    Brookfield recommends using the solvent trap at all times to enhance the temperature control of the sample material.
  • Page 11: Ii.5 Hold Time Settings

    Note: The TIMER (SEC) column will indicate the accumulated time of running at the moment the print key is pressed while the cone is rotating. This time value will not include the hold time. Brookfield Engineering Labs., Inc. Page 11 Manual No. M02-312...
  • Page 12: Ii.8 Run And Stop Keys

    Pressing and holding the STOP/ESCAPE key during power up selects the viscosity display units and speed of rotation. Pressing and holding the STOP and PRINT keys simultaneously executes the printing of a new heading (Section II.7). Brookfield Engineering Labs., Inc. Page 12 Manual No. M02-312...
  • Page 13: Operation

    See Appendix A for information on how to determine Full Scale Viscosity Range (FSR). Please contact Brookfield for accuracy information for CAP spindles 07-10. Table III-1 Accuracy of Viscosity Measurement...
  • Page 14: Iii.2 Repeatability

    Additional weights should only be considered when definitely required and removed when not required. Contact Brookfield Engineering Laboratories or your Brookfield Dealer/Distributor/Represen- tative for details on the above information.
  • Page 15 2. When measuring volatile samples such as paints and coatings, and when using either a high temperature CAP 1000+H or CAP 2000+H, the sol- vent trap must be put in place over the cone to prevent the test sample from drying out during the rotation of the cone.
  • Page 16 10. Relocate the solvent trap onto the cone adapter and raise the handle. 11. It is recommended to remove the cone for cleaning. However, with care, the cone can be cleaned in place. 12. Clean the viscometer plate (refer to Section I.8). Brookfield Engineering Labs., Inc. Page 16 Manual No. M02-312...
  • Page 17: Appendix A - Cone Numbers, Sample Sizes, Viscosity Ranges

    3.47 3,150 2.399 1.67 16.7 1.39 13.9 12,500 1.511 6.67 66.7 5.56 55.6 50,000 0.953 0.667 - 6.67 0.556 5.56 5,000 1.511 Table A-2 Number is marked on the cone shaft. Brookfield Engineering Labs., Inc. Page 17 Manual No. M02-312...
  • Page 18 Shear Rate Equation A4 sin θ 2π ω Cone Speed (rad/sec) = θ Cone Angle (degrees) radius of cone spindle Viscosity (poise) Full Scale Viscosity Range x (% FSR) Equation A5 Brookfield Engineering Labs., Inc. Page 18 Manual No. M02-312...
  • Page 19 CAP 1000+L Viscometer; Cone (02); running at 750 RPM; XAMPLE temperature 25.0°C; % FSR = 73.8. Determine the viscosity (poise), shear stress (dynes/cm ), shear rate (sec Using Equation A2, determine the full scale viscosity range at 750 RPM: Full Scale Viscosity Range (poise) = 3,750 = 5.00 poise Using Equation A5, determine the viscosity at 73.8% of full scale range:...
  • Page 20: Appendix B - Calibration Procedures

    Example: Calculate the acceptable range of viscosity using CAP 1000+ with Spindle 03 at ° 900 RPM; Brookfield Standard Fluid CAP-3L with a viscosity of 350 cP at 25 1) Calculate full scale viscosity range using the equation: Range Coefficient...