4. Preparation for the unit
(1) Place the cap(M), cap(S) for the screw, the cap (L) for the door lead wire outlet, inner door
hinge, inner door latch support and the screw to their each opposite side.
5. Inner, outer door installation
(1) Place the door to the unit, connect lead wire connecter (2pcs), ground and fix with nylon clip.
Note
Ensure the nylon clip position not to be lead wire pulled out when the door shut.
Note
Outer door screw fastening torque = more than 80kgf/cm
(2) Fit the bushing to the lead wire outlet to place the power panel back.
At the time ensure sampling air hose should not be bent.
Note
(3) Install the inner door upside down on the unit.
(4) At that time the inner door caution label is upside down.
Replace with the new caution label which is enclosed with the unit.
Inner door latch support
Cap (M)
Cap (S)
Cap (L)
Lead wire
connector
45
Ground
Details of B
Screw
Inner door
hinge
Inner door
hinge pin
Bushing for lead
wire outlet
Nylon clip