Sanyo MCO-5M Service Manual page 25

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

F29: Displays a coefficient for O
F30: Displays a ROM version
<Direction for
use>
12.
Bottom heater control
In order to shorten the recovery time of humidity, conduct electricity to humidity heater as
follows.
In the ordinary, bottom heater should be conducted maximum 80% of electricity.
When the internal temperature is SV-0.4~-0.6deg, conduct 80% electricity to humidifying
heater for 10minutes or less. In a temperature except for above, SV in F06 is applied.
In the following time period, 80% electricity conduct should not be performed:
1) The period until an internal temperature reaches to SV-0.2deg after the power is
supplied
2) The period until an internal temperature reaches to SV-0.2deg after the door is opened
continuously for 60seconds or more.
13.
Door heater control
Door heater is controlled in the ratio that was set in F07. However, door heater is controlled in
70% of main heater SV until it has controlled by PID to prevent temperature inside outer door
rising during pull-up or door opening.
14.
Remote alarm
In normal condition ......... remote alarm contact is "open"
In alarming condition ...... remote alarm contact is "close"
Note) The recovery time (after BZ key is pressed) and linkage remote alarm with buzzer
depend on F24, F25 setting.
15.
Ring Back mode
To prevent someone pressing BZ key to be buzzer alarm and remote alarm inoperative.
1. If the buzzer is cancelled by pressing BZ key, buzzer should come back after Ring Back
setting time has passed.
2. Remote alarm should not be cancelled by pressing BZ key.
3. Alarm on control panel is kept displaying.
<Alarming actions>
Status
Alarming
Normal after alarming
Alarming after setting time
has passed
Input "F29" on temperature display and press ENT key to show a coefficient
for O
span point adjustment. Press ENT key or leave it for 90seconds to
2
revert to PV display.
Input "F30" on temperature display and press ENT key to show a current
ROM version on CO
display.
2
Auto recovery 30min. setting
(F25 = 030)
Press BZ key to stop buzzer sounding
Remote alarm on control panel is kept activating
Alarm stands-by
When alarm comes back, buzzer, remote alarm and alarm on
control panel are operative.
Buzzer sounds after 30min. has
passed
since
pressed.
span point adjustment (high density) (initial: 100)
2
Auto recovery is not functioned
Buzzer does not sound when an
BZ
key
was
alarm keeps activating, even if BZ
key is pressed. At the time remote
alarm and alarm on control panel
are operative.
22
(F25 = 000)

Hide quick links: