Sanyo MCO-5M Service Manual page 23

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

Initial values in non-volatile memory are follow:
Internal temperature SV
CO
level SV
2
O
level SV
2
UV lamp lighting time SV (F01)
Humidifying heater balance value (F06)
Door heater balance value (F07)
CO
zero adjustment data (F10)
2
CO
span data (F11)
2
Internal temperature zero adjustment data (F15)
O
zero adjustment data (low density) (F26)
2
O
span data (low density) (F27)
2
O
zero adjustment data (high density) (F28)
2
O
span data (high density) (F29)
2
Key Lock data
Auto Changer data (the 3
Demo Mode data (the 2
To initialize the value of CO
memory, input "F16" on temperature display and press ENT key to display "000"
on CO
display, then change the 1
2
CO
zero adjustment data (F10) : 0
2
CO
span data (F11)
2
To initialize the value of O
memory, input "F16" on temperature display and press ENT key to display "000"
on CO
display, then change the 1
2
O
zero adjustment data (low density) (F26) : 0
2
O
span data (low density)
2
O
zero adjustment data (high density) (F28) : 0
2
O
span data (high density)
2
Note)
F14: LEDs and drivers performance check
<Direction
Input "F14" on temperature display and press ENT key to flash lamp and the all
for use>
digits at 0.5seconds intervals. Press ENT key to stop the performance.
If there is no key operation for approx. 90seconds, automatically comes to end
F15: Displays a temperature zero point adjustment value (initial: 0.8 C)
<Direction for
Input "F15" on temperature display and press ENT key to display an
use>
adjustment value on CO
key or leave the device for 90seconds, automatically reverts to PV display.
(Ex.) In "37.0" is displayed on the control panel and actual temperature is "37.8", if
you perform temperature calibration (actual temperature "37.8" is input),
"+0.8" would be housed in F15. In Function Mode, input F15 and press ENT
key to display the value on CO
F21: Communication device address (ID) set (initial: 000)
<Direction for
Input "F21" on temperature display and press ENT key to show "000" on CO
use>
display. Then input the value that identifies the product (serial communication
ID Number) and press ENT key to set serial communication ID. The settable
range is 000~255. (000: communication OFF)
rd
digit in F08)
nd
digit in F08)
zero adjustment data and span data in non volatile
2
: 100
zero adjustment data and span data in non volatile
2
Service code is not initialized.
display for temperature zero point. If you press ENT
2
20
: 37.0 C
: 0.0%
: 20.0%
: 0 minutes
: 5
: 6
: 0
: 100
: 008 (0.8 C)
: 0
: 100
: 0
: 100
: Key Lock OFF
: Auto Changer OFF
: Demo Mode OFF
st
digit to "2" (002) and press ENT key.
st
digit to "3" (003) and press ENT key.
(F27) : 100
(F29) : 100
display.
2
2

Hide quick links: