Appendix 4 Overall Sensitivity

A4.1 Overall Sensitivity of SD5 (2MHz Wired Probe)

Diameter of
Target
Distance
Reflection
Reflector
(d)(mm)
Loss A(d)
(mm)
50
1.58
75
A=45.7dB@
2MHz
100
200
50
75
2.38
A=43.2dB@
2MHz
100
200
Doppler Frequency (Hz)
Two-way Attenuation
B=∑B
∑B (T: ultrasonic attenuation
phantom No. B
:dB)
a
T
6#6#6#1#
39.8
B
78.3
a
T
6#6#6#
43.3
B
74.6
a
T
6#6#5#4#
45.8
B
72.6
a
T
6#6#4#3#
51.8
B
67.6
a
T
6#6#6#2#
37.2
B
80.0
a
T
6#6#6#1#
40.7
B
78.3
a
T
6#6#6#1#
43.2
B
78.3
a
T
6#6#4#3#
49.2
B
67.6
a
200
+B
V
(r.m.s)
a
w
S
B
B
mV
w
(dB)
(dB)
0
78.3
70.06
0
74.6
69.39
0
72.6
69.72
0
67.6
54.04
0
80.0
60.47
0
78.3
58.64
0
78.3
63.61
0
67.6
52.58
Velocity of Target
- 43 -
(
)
V
r
.
m
. .
s
=
 
 
V
(r.m.s)
s
C
20
log
(
)
n
10
V
r
.
m
. .
s
n
mV
dB
31.43
6.96
29.83
7.33
29.66
7.42
29.35
5.30
31.09
5.78
30.32
5.73
30.74
6.32
29.31
5.08
(cm/s)
Overall Sensitivity
Overall Sensitivity
(S=A(d)+B+C)
dB
125.06
125.23
125.82
124.70
122.98
124.73
127.82
121.88