Error Codes - Stryker Crossfire 2 Manual

Integrated resection and energy system
Hide thumbs Also See for Crossfire 2:
Table of Contents

Advertisement

Disposable
Exceeded time
Attachment
usage
(Continued)
RF power is too
high
RF voltage is too
high
RF current is too
high
RF delivery
has exceeded
continuous limit
Low impedance
detected
Footswitch
A wireless
footswitch is not
detected
The footswitch
icon does not
appear

Error Codes

When the Crossfire 2 system encounters an error, it will display an error code on the LCD. Error
codes are grouped into general categories that share common solutions:
1
2
Error Code
Category
A##
Activation Errors
E##
System-level Errors
P##
Probe Errors
W##
Warning Errors
RF probe
communication error
EN-26
Replace probe
Check the probe for damage.
If necessary, replace probe.
Check the probe for damage.
If necessary, replace probe.
Check the probe for damage.
If necessary, replace probe.
Clear error and continue
Check the probe for damage.
If necessary, replace probe.
Disconnect the wired footswitch.
Ensure the unit is connected.
Ensure that there is no damage to the cable or
connector.
Solution
Reactivate
Reboot system
Follow instructions on LCD, or replace disposable
attachment
No action required;
informational only
Check the connection to the console.
If necessary, replace probe.