OPENCHOICE CALIBRATION
a. Press the
OK
button to reach the Menu screens.
b. Use the Up p and Down q Arrows to scroll
until you highlight "Enter BG."
c. Press the
OK
button to select this option. You
will see a number in mg/dL units.
NOTE:
The default setting is 120 mg/dL.
d. T o enter the fingerstick reading taken from your
meter, press the Up p and Down q Arrows until
you arrive at the correct reading. Enter a reading
between 40-400 mg/dL taken within the past
5 minutes. Press the
OK
button to enter
the reading.
i) Press the OK button again to confirm the reading is correct.
ii) I
f the fingerstick reading displayed is incorrect, press the C
and enter in the correct reading for calibration.
iii) If you do not press any button, the Receiver will "time out" and no
fingerstick reading information will be entered.
e. The Glucose Readings
Processing screen will
appear, followed by the
Results Added screen.
8
|
SEVEN PLUS Continuous Glucose Monitoring System Blinded Mode
f. For Start-Up Calibration, repeat these steps for the second
g. Your continuous glucose readings will be updated every 5 minutes, but
NOTE:
Receiver within 5 minutes of being taken. If greater than 5 minutes has
passed, take new fingerstick readings.
AUTO UPLOAD CALIBRATION
a. Make sure the OneTouch Ultra Meter display is off (blank screen)
b. U se the Connection Cable to first plug the cable into the connection
c. Then connect the wide end of the cable into the Receiver. Make sure
d. T hree Receiver screens will appear in this order: Connection Screen g
button
e. Once you see the Results Screen, then press the Push button on the
fingerstick reading.
you will not see them displayed in the Status Box.
OpenChoice calibration BG readings must be entered into the
before connecting. To turn the display off, press the "M" button once.
port at the bottom of the OneTouch Ultra Meter.
the raised "Push" button on the cable is facing down. After a few
seconds "PC" will appear on the OneTouch screen.
Processing Screen g Results Screen
Connection Cable to disconnect the cable from the OneTouch Ultra
Meter and the Receiver (the order that you disconnect the cable does
not matter).