8.4 Table of settable functions (Versions 3.00, 3.01, 3.02, 3.03, 4.02, 4.03)
Function (Fct.)
Enter range
1.0
OPERATION
1.1
DISPLAY
1.1.1 FCT.DISP
Options
1.1.2 UNIT.LENGTH
Options
1.1.3 UNIT. CONV.
Options
2.0
TEST
2.1
HARDWARE
2.1.1 MASTER
Special function
2.1.2 DISPLAY
Special function
2.1.3 STATUS
Options
2.2
CUR.OUTP.I
2.2.1 VALUE I
Value display
2.2.2 TEST I
Options
2.3
SW.OUTPUT
2.3.1 TEST S
Options
2.4
FIRMWARE
2.4.1 MASTER
Display
2.4.2 DISPLAY
Display
3.0
INSTALL.
3.1
BASIS.PARAM
3.1.1 TANKHEIGHT
Options, unit
Enter
3.1.2 BLOCKDIST
Enter
3.1.3 ANTENNA
Options
3.1.4 ANT.EXTENS.
Enter
3.1.5 DIST.PIECE
Enter
Installation and operating instructions BM 70 A/P
Default value
LEVEL
LEVEL
DISTANCE
CONVERSION
REFLECTION
m
Meter [m]
cm
mm
inch
Ft
PERCENT
BARGRAPH
m3
Cubic meter [m3]
l
US Gal
GB Gal
Ft3
bbl
PERCENT
BARGRAPH
USER UNIT
Module DIS
Module MW
2 mA
2 mA
4 mA
6 mA
8 mA
10 mA
12 mA
14 mA
16 mA
18 mA
20 mA
22 mA
OPEN
OPEN
CLOSED
m
m
cm
mm
inch
Ft
10.00 m* / 20.00 m *
0.50 [m] ... max. tank height
0.50 m *
0.10 [m] ... tank height
STANDARD
STANDARD *
WAVE STICK
0.00 m *
0.00 [m] ... Tank height
0 mm *
0 ... 2000 [mm]
Description
Select function of display
(value to be displayed).
Select unit for value of length to be displayed
(level/distance).
Select unit for conversion value to be displayed
("volume table").
Tests the hardware of the master.
Tests the hardware of the display.
Displays ID-Number and Status Bytes
Displays the actual value of the current output.
Output of selected value to the current output.
IMPORTANT !!! With safety inquiry because of direct
access to the current output !!!
(Exi = min. 3.6 mA)
Switching output on/off.
IMPORTANT !!! With safety inquiry because of direct
access to the switching output !!!
Displays the firmware version of the master.
Displays the firmware version of the display.
To input the tank height.
The tank height is defined as the vertical distance
between the bottom edge of the flange and the tank
bottom.
The unit entered here is also used for all other length
inputs.
To input the block distance.
The block distance marks the non-measurable range
below the bottom edge of the flange.
Selection of the antenna type.
(WAVE STICK only for BM 70 A)
To input the length of the antenna extension.
To input the length of the flange distance piece in [mm]
(high-temperature version = 120 mm).