PROBLEM
The indicator light
on the battery
charger is flashing
orange
The indicator light
on the battery
charger does not
light
60
RESOLUTION
1.
Only connect rechargeable battery
modules to the charger.
2.
Try a different rechargeable battery
module.
3.
If you are using the Y battery charger, try
the rechargeable battery module on the
other connector on the charger.
4. The room temperature may be outside
the operating range of the charger. Try
using the charger in a different location.
5.
The USB port may not be the correct
type. Try another USB port.
6. If using a USB hub, too many devices
may be connected to the hub. Try
removing some devices from the hub.
7.
If using a USB hub, it may not be the
correct type. Use a powered hub.
1.
The rechargeable battery module is not
properly connected. Check the
connection.
2.
The rechargeable battery module is
over-discharged.
3.
The rechargeable battery module is
faulty. Try a different module.
4. There is no power to the USB port. Check
the power supply.