Philips DigiTrak XT Instructions For Use Manual page 42

Holter recorder
Table of Contents

Advertisement

Table A-3. Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The DigiTrak XT recorder is intended for use in the electromagnetic environment specified
below. The customer or the user of the DigiTrak XT recorder should assure that it is used in
such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from surfaces, objects, and people.
Additional notes are on following page.
A-4
DigiTrak XT Recorder Instructions for Use
IEC 60601
Compliance Level
Test Level
3 Vrms
3 Vrms
150 kHz to 80
3 V/m
MHz
3 V/m
80 MHz to 2,5
GHz
Electromagnetic Environment:
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the DigiTrak XT,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
D = 1.2vP
D = 1.2vP 80 MHz to 800 MHz
D = 2.3vP 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and D
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol: