Only use Philips Medical Systems replacement parts with the DigiTrak XT recorder. The use
of non-approved replacement parts may result in increased Radiated Emissions or decreased
Electromagnetic Immunity of the DigiTrak XT recorder.
Table A-1. Guidance and Manufacturer's Declaration: Electromagnetic Emissions
The DigiTrak XT recorder is intended for use in the electromagnetic environment specified
in the table below. The customer or the user of the DigiTrak XT recorder should assure that
it is used in such an environment.
Emissions Test
RF Emissions CISPR 11
Harmonic Emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
A-2
DigiTrak XT Recorder Instructions for Use
Compliance
Group 1
Not Applicable
Not Applicable
Electromagnetic
Environment: guidance
The DigiTrak XT recorder uses
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any
interference in nearby
electronic equipment.