5
Maintenance

Cleaning tasks

Cleaning materials

Replacement parts

A-61096 June 1999
To keep your scanner/microimager in good operating condition, the
following tasks should be performed:
Cleaning the imaging guides (daily)
Cleaning the Document Scanning Array (daily)
Cleaning the feed and separator rollers (daily)
Cleaning the paper path (daily)
Vacuuming inside the scanner/microimager (daily)
Cleaning the exposure lamp housings (weekly)
Cleaning the mirrors (weekly)
Cleaning the cabinet (weekly)
Cleaning the film cassette(s) (monthly)
Cleaning the air filter (monthly)
For additional maintenance and feeder adjustment information, see
Kodak publication A-61604. This publication is part of the
Maintenance and Feeder Adjustment procedure video, catalog number
828 5306.
Materials needed to perform all maintenance procedures are:
A mild cleaning agent
A quality lens cleaner
A soft, clean, lint-free cloth
A soft-bristled cleaning brush
A vacuum cleaner
Staticide Wipes
Roller Cleaning Pads
Paper Path Cleaner
IMPORTANT: Do not use any unauthorized commercial cleaning
solvent in any of the cleaning procedures.
Consumable parts that can be replaced by the customer include:
Air filter
Film cassette battery
DSA lamps
Exposure lamps
Imaging guides
Instructions on how to replace these parts are included later in this
chapter.