Status Indicators - Agilent Technologies 1260 Infinity II User Manual

Infinitylab lc series fraction collector
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

4
Using the Module

Status Indicators

Status Indicators

1 The module status indicator indicates one of six possible module
62
conditions:
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example
a leak or defective internal components).
5. Resident mode (blinking) - for example during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.
1260 Infinity II Fraction Collector User Manual