CareFusion LE200 LC Original Instruction page 57

Table of Contents

Advertisement

Option Description
OP11
Scaling of the profiles in profile mode (not for
test mode)
OP12
Unit for display of speed
OP13
Unit for display of distance
OP14
Unit for angle of elevation
OP15
Subject's body weight
(default value)
OP16
Request for body weight
before manual or automatic start
OP17
Unit for energy consumption
OP18
Maximum speed in cardio mode (default value)
(this option is only available for treadmill-
ergometers, not for ladder-ergometers)
OP19
Setting of sender for POLAR W.I.N.D. system
Comment / Display
shows: Scaling possibilities
0: No scaling (standard)
1: Scaling 1...6, which is shown in the profile mode on the INDEX
display, refers to all parameters together (speed, elevation, time)
2: Scaling 1...6 refers to each parameter individually (speed, elevation,
time)
indicates: Unit of speed
... without decimal place:
3 = m/min
... with one decimal place:
0 = km/h
1 = m/s
2 = mph 23 = m/min
... with two decimal places:
20 = km/h
21 = m/s
22 = mph
 km/h, m/s, mph or m/min flashes
indicates: Unit of distance
0 = km
1 = miles
2 = m
 m, km, or miles flashes
indicates: Unit of elevation
0 = % (per cent)
1 = ° (degree)
 % or ° flashes
indicates: 10 ... 250 (estimated weight)
 weight flashes
The personal body weight is necessary for a more correct calculation
(estimation) of power and energy consumption.
0 = OFF. Request for body weight before starting a program is not
required. Calculation of energy consumption and power is based on
the body weight entered in option no. 15.
1 = ON. Input of body weight before starting a program is required.
Calculation of energy consumption and power is based on the entered
body weight.
JOUL = kJoule is the unit of energy consumption
CALO = kcal is the unit of energy consumption
reports: 0.0 ... max
for the default value for maximum speed in the cardio mode.
 adjusted unit flashes,  max. flashes
The value of max. speed in cardio mode can be changed online by
pressing
0000 0000 = all senders are accepted (can also be set with UP and
DOWN)
xxxx xxxx = only specific sender with special ID is accepted, must
be set with
9999 9999 = next sender will be accepted, saved and filtered
(can also be set with UP and DOWN)
(further settings in user OP23 and administrator OP16 are necessary)
and
cos105000-2016-00 en
page 57